Performance Verification Test
5.6

Bandwidth Test

5.6.1

Specification

Bandwidth
Amplitude
[1]
Loss
NOTE
[1]: Amplitude Loss (dB) = 20 x lg (Vrms2/Vrms1); wherein, Vrms1 is the measurement result of
amplitude effective value at 1MHz and Vrms2 is the measurement result of amplitude effective
value at full bandwidth.
5.6.2

Test Connection Diagram

5.6.3

Test Procedures

WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
Bandwidth test when the input impedance is 1 MΩ
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.5
2. Turn on Fluke 9500B and set its impedance to 1 MΩ.
3. Configure the oscilloscope:
DHO4000 Performance Verification
28
DHO4204
DHO4404
DHO4804
DHO4000
CH1

Figure 5.5 Bandwidth Test Connection Diagram

.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
200 MHz
-3 dB, all channels mode
400 MHz
-3 dB, all channels mode
-3 dB, single channel
800 MHz (50 Ω)
mode and half channels
500 MHz (1 MΩ)
mode
Fluke 9500B