Test Record From - Rigol DHO4000 Series Performance Verification Manual

Digital oscilloscope
Hide thumbs Also See for DHO4000 Series:
Table of Contents

Advertisement

Performance Verification Test
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
Impedance test when the input impedance is 50 Ω
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.1
2. Configure the oscilloscope:
a. Use the method illustrated above to turn on CH1.
b. Set the input impedance of CH1 to 50 Ω.
c. Set the vertical scale of CH1 to 100 mV/div.
3. Turn on Fluke 9500B; set its impedance to 50 Ω and select the resistance
measurement function. Read and record the resistance measured.
4. Adjust the vertical scale of CH1 to 500 mV/div; read and record the resistance
measured.
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
5.2.4

Test Record From

1 MΩ Input Impedance
Channel
CH1
CH2
CH3
CH4
50 Ω Input Impedance
Channel
CH1
CH2
DHO4000 Performance Verification
16
.
Vertical Scale
Test Result
50 mV/div
500 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
Vertical Scale
Test Result
50 mV/div
500 mV/div
50 mV/div
500 mV/div
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Limit
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
Limit
49.5 Ω to 50.5 Ω
49.5 Ω to 50.5 Ω
49.5 Ω to 50.5 Ω
49.5 Ω to 50.5 Ω
Pass/Fail
Pass/Fail