Test Connection Diagram; Test Procedures; Figure 5.5 Bandwidth Test Connection Diagram - Rigol DHO1000 Series Manual

Digital oscilloscope
Hide thumbs Also See for DHO1000 Series:
Table of Contents

Advertisement

5.6.2

Test Connection Diagram

5.6.3

Test Procedures

WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.5
2. Turn on Fluke 9500B and set its impedance to 1 MΩ.
3. Configure the oscilloscope:
a. To enable CH1, perform any of the following operations:
- Click or tap the channel status label at the bottom of the screen to enable
- Press the front-panel
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
b. Click or tap the channel status label of CH1 at the bottom of the screen. Then
the Vertical menu is displayed. Click or tap Probe to enter the Probe setting
menu. Set the probe attenuation ratio to "1X".
c. Set the vertical scale to 50 mV/div.
d. Set the vertical offset to 0 V. Then close the Vertical menu.
e. Click or tap the horizontal label at the top of the screen. Then the Horizontal
menu is displayed. In this menu, you can set the horizontal time base to 500 ns/
div.
f. Set the horizontal position to 0 s. Close the Horizontal menu.
g. Set the trigger level to 0 V.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DHO1000
CH1

Figure 5.5 Bandwidth Test Connection Diagram

.
CH1.
turn CH1 on.
Performance Verification Test
Fluke 9500B
key to enable the channel.
DHO1000 Performance Verification