Performance Verification Test
5.3.4
Test Record From
Channel
CH1
CH2
CH3
CH4
5.4
DC Gain Accuracy Test
5.4.1
Specification
DC Gain Accuracy
Specification
5.4.2
Test Connection Diagram
5.4.3
Test Procedures
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in
Figure 5.3
2. Set the impedance of Fluke 9500B to 1 MΩ.
3. Output a DC signal with +3 mVdc voltage (Vout1) via Fluke 9500B.
DHO1000 Performance Verification
16
Vertical Scale
50 mV/div
500 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
DHO1000
CH1
Figure 5.3 DC Gain Accuracy Test Connection Diagram
.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Test Result
Limit
19 pF ± 3 pF
19 pF ± 3 pF
19 pF ± 3 pF
19 pF ± 3 pF
19 pF ± 3 pF
19 pF ± 3 pF
19 pF ± 3 pF
19 pF ± 3 pF
±2% x Full Scale
Fluke 9500B
Pass/Fail