Advertisement

5. In the Measure menu, click or tap Vertical measurement item to select "Vavg". The
Vavg result is displayed in the "Result" sidebar at the right side of the screen. Read
and record the measurement result of Vavg1.
6. Set the vertical offset of the oscilloscope to 0 V.
7. Set the output of Fluke 9500B to 0 V. Read the value from the "Result" sidebar and
record the measurement result.
8. Repeat Step 2-7. Measure and record the average results of different vertical scales
and offsets according to
5.5.4

Test Record From

Channel
Setting
CH1, 20
MHz BW
CH2, 20
MHz BW
CH3, 20
MHz BW
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Test Record From
Vertical
Offset
Scale
500 μV/div
1 V
500 μV/div
0 V
500 μV/div
-1 V
200 mV/div
10 V
200 mV/div
0 V
200 mV/div
-10 V
1 V/div
20 V
1 V/div
0 V
1 V/div
-20 V
5 V/div
100 V
5 V/div
0 V
5 V/div
-100 V
500 μV/div
1 V
500 μV/div
0 V
500 μV/div
-1 V
200 mV/div
10 V
200 mV/div
0 V
200 mV/div
-10 V
1 V/div
20 V
1 V/div
0 V
1 V/div
-20 V
5 V/div
100 V
5 V/div
0 V
5 V/div
-100 V
500 μV/div
1 V
500 μV/div
0 V
500 μV/div
-1 V
200 mV/div
10 V
Performance Verification Test
.
Test Result
Min.
-1.01 V
-2.05 mV
-982.95 mV
-10.17 V
-22.00 mV
9.92 V
-20.30 V
-102.00 mV
19.69 V
-101.50 V
-502.00 mV
98.49 V
-1.01 V
-2.05 mV
-982.95 mV
-10.17 V
-22.00 mV
9.92 V
-20.30 V
-102.00 mV
19.69 V
-101.50 V
-502.00 mV
98.49 V
-1.01 V
-2.05 mV
-982.95 mV
-10.17 V
DHO1000 Performance Verification
Max.
-982.95 mV
2.05 mV
1.01 V
-9.92 V
22.00 mV
10.17 V
-19.69 V
102.00 mV
20.30 V
-98.49 V
502.00 mV
101.50 V
-982.95 mV
2.05 mV
1.01 V
-9.92 V
22.00 mV
10.17 V
-19.69 V
102.00 mV
20.30 V
-98.49 V
502.00 mV
101.50 V
-982.95 mV
2.05 mV
1.01 V
-9.92 V
21