Performance Verification Test
2. Configure the oscilloscope:
a. To enable the channel, perform any of the following operations:
- Click or tap the channel status label at the bottom of the screen to enable
- Press the front-panel
- In the Vertical menu, select the CH1 tab. Select ON for the Display item to
b. Set the vertical scale of CH1 to 50 mV/div.
3. Turn on Fluke 9500B; set its impedance to 1 MΩ and select the resistance
measurement function. Read and record the resistance measured.
4. Adjust the vertical scale of CH1 to 500 mV/div; read and record the resistance
measured.
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
5.2.4

Test Record From

Channel
CH1
CH2
CH3
CH4
5.3

Input Capacitance Test

5.3.1

Specification

Input Capacitance
Specification
DHO1000 Performance Verification
14
CH1.
turn CH1 on.
Vertical Scale
50 mV/div
500 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
19 pF ± 3 pF
key to enable the channel.
Test Result
Limit
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
0.99 MΩ to 1.01 MΩ
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Pass/Fail