9.1
Technical data when operated with CO
6800482
Operating pressure
700 hPa to 1300 hPa
range:
Measuring range:
0 to 10000 ppm
Accuracy at 25 °C
±30 ppm, +3 % of measured value
Alarm threshold A1:
1000 ppm
Alarm threshold A2:
2000 ppm
9.2
Technical data when operated with NH
6813735
Measuring range:
0 to 100 ppm NH
Resolution:
0.5 ppm
Accuracy:
1.5 ppm or ± 10 % of measured value
(larger value applies)
Response time T
:
< 30 seconds
50
Long-term drift:
< ±10 % of measured value/year
9.3

Technical data of ambient sensors

Ambient pressure:
Temperature:
Humidity:
78
sensor
2
AL sensor
3
3
Accuracy
700 hPa to 1100 hPa = ± 1.0 hPa,
1100 hPa to 1300 hPa = ± 1.5 hPa
Accuracy
0 °C to 50 °C = ± 1.0 °C
-20 °C to 0 °C = ± 1.25 °C
5 % to 95 % RH: Accuracy ± 5 % RH
10
Order list
Article
Dräger X-node
incl. sensor (configurator)
Dräger X-node
(without sensor)
X-node app
Accessories
USB charging cable
Calibration adapter V
Mounting bracket
Weather protection
Sensors
Measuring range
DrägerSensor CO
I
0 ... 10000 ppm
2
R
DrägerSensor Cl
0 ... 100 ppm
2
DrägerSensor CO
0 ... 1000 ppm
DrägerSensor H
0 ... 3000 ppm
2
DrägerSensor H
O
0 ... 7000 ppm
2
2
HC
DrägerSensor H
O
0 ... 300 ppm
2
2
LC
Technical manual
Part number
3723669
5601008
Freely available from
Google Play Store and Apple
App Store
5610972
6810536
Not yet available
Part number
6800482
6809665
6809605
6809685
6809675
6809705
|
Dräger X-node