CEM COMPATIBILITY

EMC Directive (2004/108/EC)
All models of FCX series transmitters type FCX-AII are in accordance with :
• the harmonized standards:
- EN 61326-1 : 2006 (Electrical equipment for measurement, control and laboratory use - EMC requirements).
criteria for tranducers with integrated or remote signal conditioning)
Emission limits : EN 61326-1 : 2006
Frequency range (MHz)
30 to 230
230 to 1000
Immunity requirements : EN 61326-1 : 2006 (Table 2)
Phenomenon
Electrostatic discharge (EDS)
Rated power frequency
Burst
Surge
Conducted RF
Performance criteria :
A
B : During testing, temporary degradation or loss of function or performance which is self-recovering.
INF-TN5FCXA2V5e-E
Limits
40 dB (μV/m) quasi peack, measured at 10m distance
47 dB (μV/m) quasi peack, measured at 10m distance
Test value
4 kV (Contact)
8 kV (Air)
3 V/m (1.4 to 2.0 GHz)
1 V/m (2.0 to 2.7 GHz)
30 A/m
2 kV (5/50 NS, 5 kHz
1 kV Line to line
2 kV Line to line
3 V (150 kHz to 80 MHz)
EN 55011 / CISPR 11
Basic standard
Performance criteria
EN 61000-4-2
IEC 61000-4-2
IEC 61000-4-3
EN 61000-4-8
EN 61000-4-4
IEC 61000-4-4
EN 61000-4-5
IEC61000-4-5
EN 61000-4-6
IEC61000-4-6
Basic standard
Group 1 Class A
B
A
A
B
B
A
5