Hard Keys And Rotary Knob - Keysight Technologies B2961B User Manual

Low noise power source
Hide thumbs Also See for B2961B:
Table of Contents

Advertisement

Front Panel Reference

Hard Keys and Rotary Knob

Hard Keys and Rotary Knob
Standby switch
Turns the instrument on or off.
Trigger
Starts a single (one shot) output/measurement or initiates trigger system. If a
repeat measurement is in progress, stops it.
The single measurement is performed with the DC bias output, pulsed bias output,
arbitrary waveform output, staircase sweep output, or pulsed sweep output set to
the channel. One single measurement can contain the maximum of 100000
measurement points.
Once a single measurement starts, the data buffer (maximum 100000 data) is
cleared, and the last single measurement result is stored in the buffer. The
measurement result is displayed on the Single view, Dual view, or Graph view. Also
the last single measurement result can be listed and displayed on the Measure
Result dialog box.
Auto
Starts a repeat (continuous) measurement. If a repeat measurement is in progress,
stops it.
The repeat measurement is performed with the DC bias output of the Source
value. And the measurement result is displayed on the Single view or Dual view.
The repeat measurement result is not stored in the buffer.
The repeat measurement is performed with the following trigger setup. Settings of
"Trigger Parameters"
106
are ignored.
Acquire trigger (measurement trigger): Initiate
ARM acquire count: Infinite
ARM acquire source: AUTO (automatic internal)
TRIGger acquire count (Measure Count) 100
TRIGger acquire source (Measure Trigger): AUTO (automatic internal)
TRIGger acquire timer period (Measure Period): 10 ms (this is automatically
extended if the measurement is not completed.)
Trigger delay (Source Delay = Measure Delay): 0 s
Trigger output: Disable
Keysight B2961B/B2962B User's Guide, Edition 1