Examination And Maintenance; Figure 35: Troubleshooting Procedure (1) - Owon SDS Series Service Manual

Smart digital storage oscilloscopes
Hide thumbs Also See for SDS Series:
Table of Contents

Advertisement

5.6 Examination and Maintenance

Power on the
oscilloscope.
No
Does the display
backlight light up?
Yes
No
Does the scope
display normally?
Yes
Does the
No
oscilloscope
operate OK?
Yes
Done.
Section 5: Maintenance
Proceed to Figure 42 .
Dose the test signal of
1KHz/5V output
normally?
No
Replace
Main board.
Replace
Main board.

Figure 35: Troubleshooting procedure (1)

34
Yes
Are the Main board and
LCD connected correctly?
No
Reconnect loose
cables.
Yes
Replace the
LCD.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sds6062vSds7102vSds8102vSds8202vSds8302Sds9302

Table of Contents