Response Data Bytes
Byte
0-3
Unsigned-32
4-6
Date
7
Unsigned-8
8
Unsigned-8
9
Unsigned-8
10
Unsigned-8
Command-Specific Response Codes
Code
0
1-127
10.18 COMMAND #163: GET FAST CHANGING INFORMATION
This command is used to collect data from the Observer-i, which is likely to change at a very fast rate.
Request Data Bytes
Byte
None
N/A
Response Data Bytes
Byte
0 – 1
Unsigned-16
2 – 3
Unsigned-16
4 – 7
Float
8 – 9
Unsigned-16
10 – 11
Bit map
12
Unsigned-8
13
Unsigned-8
14
Unsigned-8
15
Unsigned-8
16
Unsigned-8
17
Integer-8
Format
Event Running Time (in Seconds)
Event Date: Day, Month, Year – 1900
Event Hour
Event Minute
Event Second
Code: 4 - Cal. OK, 5 – Cal. Adjusted, 6 – Cal. Error.
Class
Success
No Command-Specific Errors
Undefined
Format
N/A
Format
Mode – Depends on Instrument
Sub Mode – Depends on Instrument
Analog Output
Priority Fault (See Table 3: Error/Status Information)
Error Status (in word)
Alarm Hi Trigger Level
N/A
N/A
Power Cycled flag
Event Happened flag
Decibel (dB)
Description
Description
Description
Description