Command #146: Read Alarm Event Log - MSA General Monitors OBSERVER-i User Manual

Ultrasonic gas leak detector
Hide thumbs Also See for General Monitors OBSERVER-i:
Table of Contents

Advertisement

Response Data Bytes
Byte
None
N/A
Command-Specific Response Codes
Code
Class
0
Success
1-127

10.8 COMMAND #146: READ ALARM EVENT LOG

This command reads the Alarm Event Log as specified by the event log number. Event 0 is the most
recent event. Event 1 is the one just before that and so forth.
Request Data Bytes
Byte
None
N/A
Response Data Bytes
Byte
0 – 3
Unsigned-32
4– 6
Date
7
Unsigned-8
8
Unsigned-8
9
Unsigned-8
10
Unsigned-8
11
Unsigned-8
12-13
Command-Specific Response Codes
Code
0
Success
1-127
Format
N/A
Description
No Command-Specific Errors
Undefined
Format
N/A
Format
Event Running Time (in Seconds)
Event Date: Day, Month, Year – 1900
Event Hour
Event Minute
Event Second
Alarm Detected by:.
0: by Classic Mode
1: by Enhanced Mode
Peak dB during the Alarm
Undefined
Class
No Command-Specific Errors
Undefined
Description
Description
Description
Description
15
OBSERVER-i HART

Advertisement

Table of Contents