7.0 Maintenance and servicing

7.1

Basic information

Carry out a visual inspection of the unit prior to each
use including hoses, collection jar and connection ca-
ble. Damaged cables and hoses must be replaced
immediately.
Maintenance or opening and repair of the ATMOS LC
16 (with the exception of the cleaning work described
in these operating instructions) may only be carried out
by ATMOS or a specialist authorised by ATMOS. In this
case, attention should be paid to the protective technical
and hygiene measures, the notes on safety plus the
descriptions in the servicing instructions for the ATMOS
LC 16.
For repair, this device can be returned to ATMOS.
Before returning the device for repair, clean and disinfect
it.
ATMOS cannot guarantee perfect functioning neither will
it be liable for damage to people or property if:
• Any non-original ATMOS parts are used,
• the user instructions given in this manual are not
followed exactly or are disregarded,
• assembly, resetting, alterations, extensions and
repairs are not carried out by people authorised by
ATMOS.
No warranty rights shall exist in the event of damage or
failure caused by the use of non-ATMOS accessories or
non-ATMOS consumables.
Pay attention to regulations and instructions valid for the
respective application range.
14
7.2
Change of the disposable bacterial
filter plate
Remove silicone hoses on collection jar lid.
Open bacterial filter cover by turning it to the left.
Simultaneously hold silicone ring and attachment of
collection jar lid.
Lift out the disposable bacterial filter plate and replace
it by a new one.
The smooth side of the disposable bacterial filter
plate must always show in direction of the collec
tion jar. The blue mark must show up.
Please make sure that one end of the silicone
hose is connected to the attachment on the unit
and the other end to the filter attachment
(
, fig. 1, page 4) (green rings).
Attention: Failure of the unit will occur if the silico-
ne hose is connected to the attachment for suction
(
hose
, fig. 1, page 4).