Konica Minolta AeroDR SYSTEM Operation Manual page 73

Direct digitizer
Hide thumbs Also See for AeroDR SYSTEM:
Table of Contents

Advertisement

 High-dose exposure
• When high-dose exposures are performed con-
tinuously, afterimages of the last exposure will
occasionally be visible. The affect of afterimages
in this device can be resolved by leaving longer
intervals between exposures, so make the inter-
vals between exposures longer than usual when
performing high-dose exposure.
• During high-dose exposure, continual use of a
lead or other marker at the same position can
cause burning in of the afterimage. Therefore,
avoid using it in the same position. Also, if image
burn-in is visible, run calibration.
 Exposure time for basic connection/S-SRM
connection
• The maximum exposure time is as follows.
DR Detector
maximum exposure time
AeroDR 1417HQ,
AeroDR 1417S,
AeroDR 1717HQ,
0.7, 1.7, 3.2, 4.0, 6.7, 10.3
AeroDR 1012HQ,
seconds
AeroDR 2 1417HQ,
AeroDR 2 1417S
AeroDR 3 1417HD,
AeroDR 3 1717HD,
0.7, 1.7, 4.0, 10.3 seconds
AeroDR 3 1012HQ
• The exposure time for actual exposures should
not exceed the maximum exposure time you have
set. Otherwise, exposure may fail or correct im-
ages may not be able to be obtained.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• You cannot select 4.0, 6.7 and 10.3 as the maximum
exposure time for the following DR Detectors.
DR Detector
First four digits
AeroDR 1417HQ A5DP
AeroDR 1417S
A50D
AeroDR 1717HQ A6C3
AeroDR 1012HQ A5TE
• If you increase the maximum exposure time in AeroDR
1417HQ, AeroDR 1417S, AeroDR 1717HQ, AeroDR
1012HQ, AeroDR 2 1417HQ and AeroDR 2 1417S,
the DR Detector may take longer to exit sleep mode.
• When the maximum exposure time of the AeroDR 3
1417HD, AeroDR 3 1717HD and AeroDR 3 1012HQ
are set at 10.3, the interval between exposures should
be 2 minutes or longer. The following exposures also
require 2-minute intervals.
•• • •• •• •• • •••••••••• ••••••••••••••••••••••••••••••••
Serial Number
Number
50197 or less
51000 or less
51000 or less
51000 or less
73
3.2 Operation of DR Detector
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
HINT
• Use the image processing controller to set the maxi-
mum exposure time. In some cases, the maximum
exposure time is fixed according to the X-ray device.
For details, contact Konica Minolta technical represen-
tatives.
• When using ImagePilot as the image processing con-
troller, it is not possible to change the maximum expo-
sure time.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •
 Exposure time for Aero Sync mode of the
AeroDR 1417HQ, AeroDR 1417S, Aero-
DR 1717HQ, AeroDR 1012HQ, AeroDR 2
1417HQ and AeroDR 2 1417S
• Set the maximum exposure time to 0.8 seconds
or less.
• Set the minimum exposure time as follows:
DR Detector
AeroDR 1417HQ,
AeroDR 1717HQ,
AeroDR 1012HQ
AeroDR 2 1417HQ,
AeroDR 2 1417S
IMPORTANT
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
• If the conditions are not met, X-rays may not be de-
tected or streaks may occur in the image.
• Because the minimum exposure time for AeroDR
1417HQ, AeroDR 1717HQ, and AeroDR 1012HQ var-
ies depending on the X-ray device to be used, follow
the survey results at the time of installation.
• The minimum exposure time applies to the time when
X-ray is exposed onto the center of the DR Detector.
If X-ray cannot be exposed onto the center, contact
Konica Minolta technical representatives.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• • • • •
HINT
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
• When using ImagePilot as the image processing con-
troller, it is not possible to change the maximum expo-
sure time.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •
Minimum
exposure time
2.1 milliseconds
1 millisecond

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aerodr system 2Skr 3000

Table of Contents