5 .2 .3 AeroDR Interface Unit2

Status
Power LED (green) does not
light.
Feeding1/2 LED (blue) does not
light.
Communications are down
between devices, and power
LED (green) is off.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller, and
feeding1/2 LEDs (blue) are off.
Battery is not charging while the
DR Detector is used with wired
connection (the battery level on
the battery level display of the
image processing controller does
not increase), and feeding1/2
LEDs (blue) are off.

5 .2 .4 Detector Interface Unit

Status
LED(green/blue) does not light.
LED (blue) does not light.
The DR Detector is being used
with wired connection, but it is
not recognized on the image pro-
cessing controller.
Error description
AC power not supplied.
Feeding1/2 LED (blue) does not light even
when connecting the DR Detector.
The power switch is not on. Or, the power
cable is not connected.
The I/F Cable is not connected.
Error description
AC power not supplied.
The I/F cable is not connected to the DR
Detector.
The I/F cable is not connected to the DR
Detector. Or, the Ethernet cable is not
connected to the Power Supply Unit or the
hub.
111
5.2 Various problems and countermeasures
Corrective actions
Make sure the power cable is connected cor-
rectly.
Make sure that the I/F Cable is connected
correctly.
Make sure the power switch of the AeroDR
Interface Unit2 is on. Or, make sure the power
cable is connected correctly.
Make sure that the I/F Cable is connected to
the DR Detector properly.
Corrective actions
Make sure that the Ethernet cable is
connected to the Power Supply Unit properly.
Or, make sure that the DI Unit AC adapter is
connected correctly.
Make sure that the DR Detector is connected
to the I/F cable correctly.
Make sure that the DR Detector is connected
to the I/F cable correctly. Or, make sure that
the Ethernet cable is connected correctly.