 Exiting Aero Storage Mode
• Insert DR Detector into the AeroDR Battery Char-
ger and AeroDR Battery Charger2, or connect it
to a wired cable.
• Aero Storage mode exits, and DR Detector is
connected to the image processing controller.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• You can also exit Aero Storage mode by pressing and
holding the mode LED for three seconds.
•• • •• •• • •• •• •••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••
HINT
• Aero Storage mode automatically exits in the following
conditions:
– DR Detector battery level drops to less than 3 %
level.
– Aero Storage reaches the number of saved image
limit of 100 images.
– DR Detector detects an error.
•• • •• •• • •• •• •••••••••••••••••••••••••••••••••••••••••
 Exposure
• Exposure with Aero Storage is performed with the
following procedure.
•••••••••••••••••••••••••••••••••••••
IMPORTANT
• If you do not plan to expose immediately, exit Aero
Storage mode.
• In Aero Storage, the maximum exposure time is set
to 1.8 seconds and the resolution is set to the default
(200μm). Please contact Konica Minolta technical rep-
resentatives for more information about how to change
the settings.
•• • •• •• • •• •• •••••••••••••••••••••••••••••••••••••••••
1
Start Aero Storage mode .
2
Check that this device is ready to expose
images, and then prepare for the expo-
sure .
3
Push the exposure switch of the X-ray de-
vice to perform the exposure .
• After exposure, images are saved to DR Detec-
tor, and the information LED displays the num-
ber of saved images.
3.2 Operation of DR Detector
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
IMPORTANT
• The DR Detector is precision equipment, and there-
fore impact or vibration during radiography or image
transfer may affect the image quality. Be careful when
handling the DR Detector during and just after radiog-
raphy.
• The number of saved images displayed in the infor-
mation LED is counted when the orange status LED
blinks slowly after the X-ray exposure.
• The number of saved images displayed after an expo-
sure is the patient's image number.
• Images exposed using Aero Storage do not have
patient information. It is recommended, therefore, to
keep record so that you can associate the patient ex-
posed with the number of saved images displayed on
DR Detector.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •
4
Exit Aero Storage mode .
• Insert DR Detector into the AeroDR Battery
Charger and AeroDR Battery Charger2, or con-
nect it to a wired cable.
• Aero Storage mode exits, and DR Detector is
connected to the image processing controller.
5
Import the images using the image pro-
cessing controller .
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
IMPORTANT
• While importing the images, do not disconnect DR De-
tector from AeroDR Battery Charger, AeroDR Battery
Charger2, or the wired cable.
• When the image import process finishes, the number
of saved image in DR Detector becomes zero.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •