7.1 Specifications

7 .1 .18 SKR 3000

Item
When
operating
Recommend-
ed storage
and usage
environment
conditions
When not
operating
In storage/
transport
Safety IEC60601-1 Class I
Classification
Operation
Continuous operation
mode
Temperature
35 °C
10 °C to 35 °C
10 °C
* The temperature and humidity when operating the optional components, except Detector Inter-
face Unit and Power Supply Unit, are: 10 °C to 30 °C and 35 % to 80 %, respectively.
* For serial exposure, temperature: 10 °C to 30 °C, humidity 35 % to 85 %.
* The temperature and humidity when operating the Detector Interface Unit 2, Generator Interface
Unit 3, I/F Cable4 8mD and I/F Cable4 8mU, are: 10 °C to 30 °C and 35 % to 85 %, respectively.
* Limit continuous use in a hot and humid environment (35 °C to 37 °C/95 % or lower) of an incu-
bator to within 25 minutes.
Temperature
40 °C
–10 °C to 40 °C
–10 °C
Temperature
50 °C
*1
–20 °C to 50 °C
-20 °C
*1 However, performance warranty period when storing at 50 °C is 6 months after packing.
Description
Humidity
35 %RH to 85 %RH
(ensure no water
condensation)
35 %RH
Humidity
20 %RH to 90 %RH
(ensure no water
condensation)
20 %RH
Humidity
20 %RH to 90 %RH
(ensure no water
condensation)
20 %RH
134
Atmospheric pressure
85 %RH
700 hPa to 1060 hPa
Atmospheric pressure
90 %RH
700 hPa to 1060 hPa
Atmospheric pressure
90 %RH
700 hPa to 1060 hPa
1060 hPa
700 hPa
1060 hPa
700 hPa
1060 hPa
700 hPa