wtw Cond 3300i Operating Manual page 21

Handheld meter
Hide thumbs Also See for Cond 3300i:
Table of Contents

Advertisement

Cond 3300i/3400i
ba75684e02
03/2008
6
Press the <RUN/ENTER> key.
– If no temperature sensor is connected, enter the current
temperature of the solution with <▲> <▼>
and confirm with <RUN/ENTER>.
– If a temperature sensor is connected, the AR measurement to
determine the cell constant starts.
The AR display indicator flashes until a stable signal is reached. The
cell constant determined is displayed. The measuring instrument au-
tomatically stores the cell constant.
UpH
O
TDS
S
1
C
A L
0 7
4
Time
Baud
Day.Month No.
Year
Ident
Tref20
Tref25
AutoCal DIN
LoBat
nLF
AutoCal TEC
Lin
Cal
Note
If the error message E3 appears, refer to chapter 6 W
7
Switch to the measuring mode with <M> or <RUN/ENTER> .
Sal
mV/pH
% mg/l
m
S/
cm
mbar
M
cm
1/
cm
Sal
6
°
C
/
%
K
TP
°
F
Auto Store
ARng
AR
RCL
Operation
...
HAT TO DO IF