Cond 3300i/3400i
Calibration
evaluation
ba75684e02
03/2008
Sample printout:
CALIBRATION PROTOCOL
14.04.01
11:37
Device No.: 99990000
Calibration Conductivity
Cal time: 14.04.01 / 11:37
Cal interval:
180d
Cal Std.:
0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const :
0.478
Probe :
+++
After the calibration, the measuring instrument automatically evaluates the
current status of the calibration. The evaluation appears on the display.
Display
E3
Eliminate the error according to
chapter 6 W
HAT TO DO IF
1/cm
Cell constant [cm
0.450 ... 0.500 cm
0.800 ... 1.200 cm
outside the ranges
0.450 ... 0.500 cm
or
...
0.800 ... 1.200 cm
Operation
-1
]
-1
-1
-1
-1