wtw Cond 3300i Operating Manual page 17

Handheld meter
Hide thumbs Also See for Cond 3300i:
Table of Contents

Advertisement

Cond 3300i/3400i
AutoRead AR
(drift control)
ba75684e02
03/2008
5
Wait for a stable measured value.
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable effect on the reproducibility of the
measured value.
1
Select the required measuring mode with <M>.
2
Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
TDS
UpH
O
S
1
5
3 0
8
2
5
Time
Baud
Day.Month No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
4
If necessary, start the next AutoRead measurement with <RUN/EN-
TER>.
5
To terminate AutoRead: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing <RUN/ENTER>.