Proline Promass F 200
5.
Insert dummy plugs into unused cable entries.
7.5
Post-connection check
Are cables or the device undamaged (visual inspection)?
Do the cables comply with the requirements (→ 23)?
Do the cables have adequate strain relief?
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap"
(→ 32) ?
Depending on the device version: are all the device plugs firmly tightened (→ 29)?
Does the supply voltage match the specifications on the transmitter nameplate (→ 27)?
Is the terminal assignment correct ?
Is the terminal assignment or the pin assignment of the device plug (→ 25) correct?
If supply voltage is present, do values appear on the display module?
Are all housing covers installed and firmly tightened?
Is the securing clamp tightened correctly?
Endress+Hauser
Electrical connection
A0013960
33