Proline Promass F 200
8 mm
6.2.5
Turning the display module
6.3

Post-installation check

Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure (refer to the chapter on "Material load curves" of the "Technical Information"
document on the CD-ROM provided)
• Ambient temperature (→  17)
• Measuring range
Has the correct orientation for the sensor been selected ?
• According to sensor type
• According to medium temperature
• According to medium properties (outgassing, with entrained solids)
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the piping
(→  16)?
Are the measuring point identification and labeling correct (visual inspection)?
Endress+Hauser
max. 350°
1
3 mm
Installation
8 mm
A0013713
A0013905
21