Endress+Hauser Proline Promass F 200 Brief Operating Instructions page 28

Coriolis flowmeter
Hide thumbs Also See for Proline Promass F 200:
Table of Contents

Advertisement

Electrical connection
An external power supply is required for each output. The following supply voltage values
apply for PROFIBUS PA and the pulse/frequency/switch output:
Order code for "Output"
1) 2)
Option A
: 4-20 mA HART
1) 2)
Option B
: 4-20 mA HART, pulse/frequency/switch
output
1) 2)
Option C
: 4-20 mA HART, 4-20 mA
3)
Option G
: PROFIBUS PA, pulse/frequency/switch output
1)
External supply voltage of the power supply unit with load.
2)
For device versions with SD03 local display: The terminal voltage must be increased by DC 2 V if backlighting is
used.
3)
For device version with SD03 local display: The terminal voltage must be increased by DC 0.5 V if backlighting is
used.
Load
Load for current output: 0 to 500 Ω, depending on the external supply voltage of the power
supply unit
Calculation of the maximum load
Depending on the supply voltage of the power supply unit (U
including line resistance must be observed to ensure adequate terminal voltage at the device.
In doing so, observe the minimum terminal voltage (→  27)
• For U
= 18 to 18.9 V: R
S
• For U
= 18.9 to 24.5 V: R
S
• For U
= 24.5 to 30 V: R
S
28
≤ (U
- 18 V) :0.0036 A
B
S
≤ (U
- 13.5 V) :0.022 A
B
S
≤500 Ω
B
Minimum terminal
voltage
For 4 mA: ≥ DC 18 V
For 20 mA: ≥ DC 14 V
For 4 mA: ≥ DC 18 V
For 20 mA: ≥ DC 14 V
For 4 mA: ≥ DC 18 V
For 20 mA: ≥ DC 14 V
9 V
), the maximum load (R
S
Proline Promass F 200
Maximum terminal
voltage
DC 35 V
DC 35 V
DC 30 V
DC 32 V
)
B
Endress+Hauser

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Promass F 200 and is the answer not in the manual?

Questions and answers

Table of Contents