Summary of Contents for Endress+Hauser Proline Prowirl F 200
Page 1
Products Solutions Services BA01222D/06/EN/01.13 71225412 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Prowirl F 200 PROFIBUS PA Vortex flowmeter...
Page 2
• The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these Instructions. Endress+Hauser...
Proline Prowirl F 200 PROFIBUS PA Table of contents Table of contents 6.2.5 Mounting the transmitter of the Document information ....6 remote version ....26 Document function .
Document information Proline Prowirl F 200 PROFIBUS PA Document information Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
Proline Prowirl F 200 PROFIBUS PA Document information 1.2.3 Tool symbols Symbol Meaning Flat blade screwdriver A0011220 Allen key A0011221 Open-ended wrench A0011222 1.2.4 Symbols for certain types of information Symbol Meaning Allowed Indicates procedures, processes or actions that are allowed.
The following document types are available: • On the CD-ROM supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download For a detailed list of the individual documents along with the documentation code 1.3.1...
Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
IT security measures in line with operators' security standards and designed to provide additional protection for the device and device data transfer must be implemented by the operators themselves. Endress+Hauser can be contacted to provide support in performing this task. Endress+Hauser...
Incoming acceptance and product identification Proline Prowirl F 200 PROFIBUS PA Incoming acceptance and product identification Incoming acceptance A0015502 A0013843 Is the order code on the delivery note (1) identical to the order code on the product sticker (2)? A0013695...
Proline Prowirl F 200 PROFIBUS PA Incoming acceptance and product identification If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. Product identification The following options are available for identification of the measuring device: • Nameplate specifications •...
Incoming acceptance and product identification Proline Prowirl F 200 PROFIBUS PA 4.2.2 Sensor nameplate Order code for "Housing" option B "GT18 two-chamber, 316L" and option K "GT18 two-chamber, remote, 316L" Ser. no.: Size: Qmax(G): Ptest: Materials: Gasket: A0020760 3...
Page 15
Proline Prowirl F 200 PROFIBUS PA Incoming acceptance and product identification Order code for "Housing" option C "GT20 two-chamber, aluminum coated" Ser. no.: Size: Qmax(G): Ptest: Materials: Gasket: A0020758 4 Example of a sensor nameplate Nominal diameter of the sensor...
Incoming acceptance and product identification Proline Prowirl F 200 PROFIBUS PA Order code for "Housing" option J "GT20 two-chamber, remote, aluminum coated" Order code: Ser. no.: Ext. ord. cd.: Size: Qmax(G): Ptest: Materials: Gasket: A0020759 5 Example of a sensor nameplate...
Proline Prowirl F 200 PROFIBUS PA Storage and transport Storage and transport Storage conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections.
Storage and transport Proline Prowirl F 200 PROFIBUS PA Packaging disposal All packaging materials are environmentally friendly and 100% recyclable: • Measuring device secondary packaging: polymer stretch film that conforms to EC Directive 2002/95/EC (RoHS). • Packaging: – Wood crate, treated in accordance with ISPM 15 standard, which is confirmed by the affixed IPPC logo.
Proline Prowirl F 200 PROFIBUS PA Mounting Mounting Installation conditions No special measures such as supports are necessary. External forces are absorbed by the construction of the device. 6.1.1 Mounting position Mounting location A0015543 Orientation Vortex meters require a fully developed flow profile as a prerequisite for correct volume flow measurement.
Page 20
Mounting Proline Prowirl F 200 PROFIBUS PA Orientation Compact version Remote version 4) 5) Horizontal orientation, transmitter head down A0015590 Horizontal orientation, transmitter head at side A0015592 In the case of liquids, there should be upward flow in vertical pipes to avoid partial pipe filling (Fig. A).
Page 22
Mounting Proline Prowirl F 200 PROFIBUS PA ATEX II1/2G Ex d, Ex ia: –40 to +60 °C (–40 to +140 °F) Local display –20 to +60 °C (–4 to +140 °F) Additionally available as order code for "Test, certificate", option JN "Transmitter ambient temperature –50 °C (–58 °F)".
Page 23
Proline Prowirl F 200 PROFIBUS PA Mounting NOTICE Electronics overheating on account of thermal insulation! ‣ Observe the maximum permitted insulation height of the transmitter neck so that the transmitter head and/or the connection housing of the remote version is completely free.
If the required inlet runs cannot be observed, it is possible to install a specially designed open aperture flow conditioner which can be ordered from Endress+Hauser. The flow conditioner is fitted between two pipe flanges and centered by the mounting bolts.
Proline Prowirl F 200 PROFIBUS PA Mounting Installation for delta heat measurements Order code for "Sensor version", option 3 "Mass flow (integrated temperature measurement)" The second temperature measurement is taken using a separate temperature sensor. The measuring device reads in this value via a communication interface.
Mounting Proline Prowirl F 200 PROFIBUS PA 3. Remove stick-on label on the electronics compartment cover. 6.2.3 Mounting the measuring device WARNING Danger due to improper process sealing! ‣ Ensure that the inside diameters of the gaskets are greater than or equal to that of the process connections and piping.
Page 27
Proline Prowirl F 200 PROFIBUS PA Mounting CAUTION Excessive force can damage the housing! ‣ Avoid excessive mechanical stress. The transmitter of the remote version can be mounted in the following ways: • Wall mounting • Pipe mounting Wall mounting 19 (0.6)
Mounting Proline Prowirl F 200 PROFIBUS PA Pipe mounting 20…70 ( 0.79 to 2.75) SW 13 A0019862 9 Engineering unit mm (in) Pipe mounting set 6.2.6 Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned: max.
Proline Prowirl F 200 PROFIBUS PA Mounting 6.2.7 Turning the display module 3 mm A0013905 1. Loosen the securing clamp of the electronics compartment cover using an Allen key. 2. Unscrew cover of the electronics compartment from the transmitter housing.
Electrical connection Proline Prowirl F 200 PROFIBUS PA Electrical connection The measuring device does not have an internal circuit breaker. For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
Page 31
Proline Prowirl F 200 PROFIBUS PA Electrical connection Connecting cable (reinforced) Cable, reinforced 4 × 2 × 0.34 mm (22 AWG) PVC cable with common shield (4 pairs, pair- stranded) and additional steel-wire braided sheath Flame resistance According to DIN EN 60332-1-2...
Page 32
Electrical connection Proline Prowirl F 200 PROFIBUS PA Non-hazardous area: • Siemens 6XV1 830-5BH10 • Belden 3076F • Kerpen CEL-PE/OSCR/PVC/FRLA FB-02YS(ST)YFL Maximum overall cable length The maximum network expansion depends on the type of protection and the cable specifications. The overall cable length combines the length of the main cable and the length of all spurs >1 m (3.28 ft).
Proline Prowirl F 200 PROFIBUS PA Electrical connection 7.1.3 Terminal assignment Transmitter Connection version for PROFIBUS PA, pulse/frequency/switch output A0013570 A0018161 Maximum number of terminals Maximum number of terminals for order code for "Accessory mounted", option NA "Overvoltage protection" Output 1: PROFIBUS PA...
Electrical connection Proline Prowirl F 200 PROFIBUS PA A0019335 10 Terminals for connection compartment in the transmitter wall holder and the sensor connection housing Terminals for connecting cable Grounding via the cable strain relief Terminal number Assignment Cable color...
Page 35
Proline Prowirl F 200 PROFIBUS PA Electrical connection Where applicable, national installation regulations and guidelines must be observed during the installation! Where there are large differences in potential between the individual grounding points, only one point of the shielding is connected directly with the reference ground. In systems without potential equalization, therefore, cable shielding of fieldbus systems should only be grounded on one side, for example at the fieldbus supply unit or at safety barriers.
Electrical connection Proline Prowirl F 200 PROFIBUS PA 7.1.6 Requirements for the supply unit Supply voltage Transmitter Supply voltage for a compact version without a local display Order code for "Output" Minimum Maximum terminal voltage terminal voltage Option G: PROFIBUS PA, pulse/frequency/ ≥DC 9 V...
Page 37
Proline Prowirl F 200 PROFIBUS PA Electrical connection Connection via terminals 3 mm 20 mm 10 (0.4) mm (in) A0013836 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Electrical connection Proline Prowirl F 200 PROFIBUS PA Removing a cable 3 (0.12) mm (in) A0013835 ‣ To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between the two terminal holes while simultaneously pulling the cable end out of the terminal.
Page 39
Proline Prowirl F 200 PROFIBUS PA Electrical connection Connecting the sensor connection housing 3 mm A0020410 A0020411 1. Loosen the securing clamp. 2. Unscrew the housing cover. 3. Guide the connecting cable through the cable entry and into the connection housing (if using a connecting cable without an M12 device plug, use the shorter stripped end of the connecting cable).
Page 40
Electrical connection Proline Prowirl F 200 PROFIBUS PA Connection to the wall holder of the transmitter Connecting the transmitter via plug A0020412 ‣ Connect the plug. Connecting the transmitter via terminals 3 mm A0020404 TX 10 8 mm A0020405 Endress+Hauser...
Page 41
Proline Prowirl F 200 PROFIBUS PA Electrical connection ~15° A0020406 A0020407 A0020409 1. Loosen the securing clamp of the transmitter housing. 2. Loosen the securing clamp of the electronics compartment cover. 3. Unscrew the electronics compartment cover. 4. Pull out the display module with a gentle rotational movement. To make it easier to access the lock switch, attach the display module to the edge of the electronics compartment.
Electrical connection Proline Prowirl F 200 PROFIBUS PA 7. Disconnect the signal cable from the connection board of the wall housing by pressing in the locking clip on the connector. 8. Remove the transmitter housing. 9. Guide the connecting cable through the cable entry and into the connection housing (if using a connecting cable without an M12 device plug, use the shorter stripped end of the connecting cable).
Proline Prowirl F 200 PROFIBUS PA Electrical connection Hardware settings 7.4.1 Setting the device address PROFIBUS PA The address must always be configured for a PROFIBUS DP/PA device. The valid address range is between 1 and 126. In a PROFIBUS DP/PA network, each address can only be assigned once.
Electrical connection Proline Prowirl F 200 PROFIBUS PA A0015903 14 Example of software addressing; switch 8 is set to the "ON" position; the address is defined in the operating menu ("Setup" menu→"Communication" submenu→"Device address" parameter). Ensuring the degree of protection The measuring device fulfills all the requirements for the IP66/67 degree of protection, Type 4X enclosure.
Proline Prowirl F 200 PROFIBUS PA Operability Operability Overview of operation options A0019091 Local operation via display module Computer with operating tool (e.g. FieldCare, SIMATIC PDM) Control system (e.g. PLC) Endress+Hauser...
Operability Proline Prowirl F 200 PROFIBUS PA Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu with menus and parameters (→ 172) Operating menu for operators and maintenances...
Proline Prowirl F 200 PROFIBUS PA Operability 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles. Each user role corresponds to typical tasks within the device lifecycle. Menu User role and tasks Content/meaning...
Operability Proline Prowirl F 200 PROFIBUS PA Access to the operating menu via the local display 8.3.1 Operational display X X X X X X X 1120.50 A0016502 Operational display Device tag (→ 69) Status area Display area for measured values (4-line) Operating elements (→...
Page 49
Proline Prowirl F 200 PROFIBUS PA Operability Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured variable Measurement channel Diagnostic behavior number ↓ ↓ ↓ Example A0013962 A0013710 A0013948 Appears only if a diagnostics event is present for this measured variable.
Operability Proline Prowirl F 200 PROFIBUS PA 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Volume flow Locking status Display A0013993-EN A0016327-EN Navigation view Navigation path to current position...
Page 51
Proline Prowirl F 200 PROFIBUS PA Operability Display area Menus Symbol Meaning Operation Appears: • In the menu next to the "Operation" selection A0013973 • At the left in the navigation path in the "Operation" menu Setup Appears: • In the menu next to the "Setup" selection A0013974 •...
Operability Proline Prowirl F 200 PROFIBUS PA 8.3.3 Editing view Numeric editor Text editor A0013941 A0013999 Editing view Display area of the entered values Input mask Operating elements (→ 53) Input mask The following input symbols are available in the input mask of the numeric and text editor:...
Proline Prowirl F 200 PROFIBUS PA Operability Selection of letters from a to z. abc _ … A0019094 Selection of special characters. "'^ _ … ~& A0019095 Confirms selection. A0013985 Switches to the selection of the correction tools. A0013987 Exits the input without applying the changes.
Operability Proline Prowirl F 200 PROFIBUS PA Meaning Enter key For operational display • Pressing the key briefly opens the operating menu. • Pressing the key for 2 s opens the context menu. In a menu, submenu • Pressing the key briefly: –...
Page 55
Proline Prowirl F 200 PROFIBUS PA Operability 2. Press + simultaneously. The context menu is closed and the operational display appears. Calling up the menu via the context menu 1. Open the context menu. 2. Press to navigate to the desired menu.
Operability Proline Prowirl F 200 PROFIBUS PA 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
Proline Prowirl F 200 PROFIBUS PA Operability Navigation path "Expert" menu → Direct access The direct access code consists of a 4-digit number and the channel number, which identifies the channel of a process variable: e.g. 0914-1. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
Operability Proline Prowirl F 200 PROFIBUS PA 8.3.9 Changing the parameters For a description of the editing display - consisting of text editor and numeric editor - with symbols (→ 52), for a description of the operating elements (→ 53) Example: Changing the tag name in the "Tag description"...
Proline Prowirl F 200 PROFIBUS PA Operability 8.3.10 User roles and related access authorization The two user roles "Operator" and "Maintenance" have different write access to the parameters if the customer defines a user-specific access code. This protects the device configuration via the local display from unauthorized access (→...
Operability Proline Prowirl F 200 PROFIBUS PA Switching off the keypad lock ‣ The keypad lock is switched on. Press the + + keys simultaneously. The message Keylock off appears on the display: The keypad lock is switched off.
PROFIBUS DP network PROFIBUS PA network Measuring device T-box Via service interface (CDI) A0020545 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device Commubox FXA291 Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291" Endress+Hauser...
FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
Proline Prowirl F 200 PROFIBUS PA Operability 8.4.3 SIMATIC PDM Function scope SIMATIC PDM is a standardized, manufacturer-independent program from Siemens for the operation, configuration, maintenance and diagnosis of intelligent field devices via PROFIBUS PA protocol. Source for device description files See data (→...
System integration Proline Prowirl F 200 PROFIBUS PA System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.00 • On the title page of the Operating instructions • On transmitter nameplate • Parameter firmware version Diagnostics →...
Proline Prowirl F 200 PROFIBUS PA System integration The fact that the manufacturer-specific GSD should be used is specified in the Ident number selector parameterManufacturer option. Where to acquire the manufacturer-specific GSD: www.endress.com → Download Area 9.2.2 Profile GSD Differs in terms of the number of Analog Input blocks (AI) and the measuring principles. If a system is configured with a Profile GSD, it is possible to exchange devices made by different manufacturers.
Page 66
System integration Proline Prowirl F 200 PROFIBUS PA • If the Prowirl 200 PROFIBUS PA is acyclically configured via an operating program (Class 2 master), access is directly via the block structure or the parameters of the measuring device. • If parameters have been changed in the device to be replaced (Prowirl 72 PROFIBUS...
Proline Prowirl F 200 PROFIBUS PA Commissioning Commissioning 10.1 Function check Before commissioning the device, make sure that the post-installation and post- connection checks have been performed. • "Post-installation check" checklist (→ 29) • "Post-connection check" checklist (→ 44) 10.2...
Page 68
Commissioning Proline Prowirl F 200 PROFIBUS PA X X X X X X X 20.50 Main menu 0104-1 Language English Display/operat. Setup Language 0104-1 English à Deutsch Español Français Language 0104-1 à English Deutsch Español Français Hauptmenü 0104-1 Sprache Deutsch...
Proline Prowirl F 200 PROFIBUS PA Commissioning 10.5 Configuring the measuring device The Setup menu with its guided wizards contains all the parameters needed for standard operation. Navigation to the Setup menu X X X X X X X 20.50...
Commissioning Proline Prowirl F 200 PROFIBUS PA XXXXXXXXX A0013375 19 Header of the operational display with tag name Device tag Navigation "Setup" menu → Device tag Parameter overview with brief description Parameter Description User entry Factory setting Device tag Enter the name for the measuring point.
Page 71
Proline Prowirl F 200 PROFIBUS PA Commissioning Length unit Parameter overview with brief description Parameter Description Selection Factory setting Volume flow unit Select volume flow unit. Unit choose list Country-specific: • l/h Result • gal/min (us) The selected unit applies for: •...
Page 72
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter Description Selection Factory setting Calorific value unit Select calorific value unit. Unit choose list Country-specific: • kJ/kg • Btu/lb Velocity unit Select velocity unit. Unit choose list Country-specific: • m/s • ft/s Density unit Select density unit.
Proline Prowirl F 200 PROFIBUS PA Commissioning 10.5.4 Selecting and setting the medium The Medium selection wizard guides you systematically through all the parameters that have to be configured for selecting and setting the medium. Navigation "Setup" menu → Medium selection...
Page 74
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter Prerequsite Description Selection Factory setting Select liquid type For the following order code: Select measured liquid type. • Water Water "Sensor version", option "Mass • LPG flow" • User-specific liquid In the Select medium parameter the Liquid option must be selected.
Proline Prowirl F 200 PROFIBUS PA Commissioning 10.5.5 Configuring the analog inputs The Analog inputs submenu guides you systematically to the individual Analog input 1 to 4 submenu. From here you get to the parameters of the individual analog input.
Page 76
Commissioning Proline Prowirl F 200 PROFIBUS PA Structure of the wizard Format display Value 1 display 0% bargraph value 1 100% bargraph value 1 Value 2 display Value 3 display 0% bargraph value 3 100% bargraph value 3 Value 4 display...
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter overview with brief description Parameter Description Selection / User entry Factory setting Format display Select how measured values are shown on • 1 value, max. size 1 value, max. size the display. • 1 bargraph + 1 value •...
Commissioning Proline Prowirl F 200 PROFIBUS PA 10.5.8 Configuring the low flow cut off The Low flow cut off wizard guides you systematically through all the parameters that have to be set for configuring the low flow cut off. Navigation "Setup"...
Proline Prowirl F 200 PROFIBUS PA Commissioning 10.6 Advanced settings The Advanced setup submenu with its submenus contains parameters for specific settings. "Setup" menu → Advanced setup Navigation to the "Advanced setup" submenu X X X X X X X 20.50...
Commissioning Proline Prowirl F 200 PROFIBUS PA Totalizer 1 to 3 (→ 94) Display (→ 96) Heartbeat (→ 171) Configuration backup (→ 98) display Administration → Define access code (→ 101) Device reset (→ 141) Order code for "Application package", option EB "Heartbeat Verification", see the Special Documentation for...
Page 81
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter overview with brief description Parameter Description Selection / User entry Factory setting Enthalpy type Define which kind of enthalpy is used. • Heat Heat • Calorific value Calorific value type Select calculation based on gross calorific •...
Page 82
Commissioning Proline Prowirl F 200 PROFIBUS PA Mol% CH4 Mol% Cl2 Mol% CO Mol% CO2 Mol% H2 Mol% H2O Mol% H2S Mol% HCl Mol% He Mol% i-C4H10 Mol% i-C5H12 Mol% Kr Mol% N2 Mol% n-C10H22 Mol% n-C4H10 Mol% n-C5H12 Mol% n-C6H14...
Page 83
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter overview with brief description Parameter Description Selection / User entry Factory setting Gas type Select measured gas type. • Hydrogen H2 Methane CH4 • Helium He • Neon Ne • Argon Ar • Krypton Kr •...
Page 84
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter Description Selection / User entry Factory setting Mol% CO2 Enter amount of substance for the gas 0 to 100 % mixture. Mol% H2 Enter amount of substance for the gas 0 to 100 % mixture.
Proline Prowirl F 200 PROFIBUS PA Commissioning 10.6.2 Performing external compensation The External compensation submenu contains parameters which can be used to enter external or fixed values. These values are used for internal calculations. Navigation "Expert" menu → Sensor → External compensation...
Commissioning Proline Prowirl F 200 PROFIBUS PA Navigation "Setup" menu → Advanced setup → Sensor adjustment Structure of the submenu Sensor adjustment → Inlet configuration Inlet run Mating pipe diameter Installation factor Parameter overview with brief description Parameter Description Selection / User entry...
Page 87
Proline Prowirl F 200 PROFIBUS PA Commissioning Structure of the wizard for the pulse output Operating mode Pulse Assign pulse output Corrected Heat flow Total mass flow volume flow difference Energy flow Mass flow Volume flow Corrected Energy unit volume unit...
Page 88
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter Description Selection / User entry Factory setting Corrected volume unit Select corrected volume unit. Unit choose list Country-specific: • Nl Result • Sft³ The selected unit is taken from: Corrected volume flow unit parameter Mass unit Select mass unit.
Page 89
Proline Prowirl F 200 PROFIBUS PA Commissioning Structure of the wizard for the frequency output Operating mode Frequency Assign frequency output Temperature Steam quality Energy flow Volume flow Mass flow Corrected Calculated saturat. Total mass Heat flow Flow velocity steam pressure...
Page 90
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter overview with brief description Parameter Description Selection / User entry Factory setting Operating mode Define the output as a pulse, frequency or • Pulse Pulse switch output. • Frequency • Switch Assign frequency output Select process variable for frequency output.
Page 91
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter Description Selection / User entry Factory setting Measuring value at maximum Enter measured value for maximum Signed floating-point number frequency frequency. Failure mode Define output behavior in alarm condition. • Actual value 0 Hz •...
Page 92
Commissioning Proline Prowirl F 200 PROFIBUS PA Operating mode Switch Switch output function Limit Status Assign status Assign limit Steam Volume Energy Reynolds Low flow cut Temp. Mass flow quality flow flow number Flow Corrected Calc. sat. Total mass Heat flow...
Page 93
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter Description Selection / User entry Factory setting Assign limit Select process variable for limit function. • Volume flow Volume flow • Corrected volume flow • Mass flow • Flow velocity • Temperature • Calculated saturated steam pressure •...
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter Description Selection / User entry Factory setting Switch-off value Enter measured value for the switch-off Signed floating-point number 0 m³/h point. Switch-on delay Define delay for the switch-on of status 0.0 to 100.0 s 0.0 s...
Page 95
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter Prerequsite Description Selection Factory setting Totalizer operation mode – • Net flow total Net flow total • Forward flow total • Reverse flow total • Last valid value Failure mode – • Stop Actual value •...
Commissioning Proline Prowirl F 200 PROFIBUS PA 10.6.6 Carrying out additional display configurations In the "Display" submenu you can set all the parameters involved in the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display Structure of the submenu Display →...
Page 97
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter overview with brief description Parameter Description Selection / User entry Factory setting Format display Select how measured values are shown on • 1 value, max. size 1 value, max. size the display. • 1 bargraph + 1 value •...
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameter Description Selection / User entry Factory setting Language Set display language. • English English • Deutsch • Français • Español • Italiano • Nederlands • Portuguesa • Polski • русский язык (Russian) • Svenska •...
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter Description User interface / Selection Factory setting Configuration management Select action for managing the device data in • Cancel Cancel the display module. • Execute backup • Restore • Duplicate • Compare • Clear backup data...
Page 100
Commissioning Proline Prowirl F 200 PROFIBUS PA Switch status Simulation device alarm Diagnostic event category Simulation diagnostic event Parameter overview with brief description Parameter Prerequsite Description Selection / User Factory setting entry Assign simulation process variable – Select a process variable for •...
Proline Prowirl F 200 PROFIBUS PA Commissioning Parameter Prerequsite Description Selection / User Factory setting entry Diagnostic event category – Select the category of the • Sensor Sensor diagnostic event. • Electronics • Configuration • Process Simulation diagnostic event – Enter the diagnostic number...
Commissioning Proline Prowirl F 200 PROFIBUS PA Parameters which can always be modified via the local display Certain parameters that do not affect the measurement are excepted from write protection via the local display. Despite the defined access code, these parameters can always be modified even if the other parameters are locked.
Page 103
Proline Prowirl F 200 PROFIBUS PA Commissioning A0013909 4. Setting the write protection switch (WP) on the main electronics module to the ON position enables the hardware write protection. Setting the write protection switch (WP) on the main electronics module to the OFF position (factory setting) disables the hardware write protection.
Operation Proline Prowirl F 200 PROFIBUS PA Operation 11.1 Reading device locking status The write protection types that are currently active can be determined using the Locking status parameter. Navigation "Operation" menu → Locking status Function scope of "Locking status" parameter...
Proline Prowirl F 200 PROFIBUS PA Operation Temperature Calculated saturated steam pressure Steam quality Total mass flow Condensate mass flow Energy flow Heat flow difference Reynolds number Density Pressure Compressibility factor Parameter overview with brief description Parameter Description User interface Volume flow Displays the volume flow currently measured.
Operation Proline Prowirl F 200 PROFIBUS PA Totalizer status (Hex) Totalizer status Totalizer status (Hex) Parameter overview with brief description Parameter Prerequsite Description Selection / User Factory setting entry / User interface Assign process variable – Assignment of a process • Volume flow Volume flow variable to the totalizer.
Proline Prowirl F 200 PROFIBUS PA Operation Parameter overview with brief description Parameter Description User interface Factory setting Terminal voltage 1 Displays the current terminal voltage that is 0.0 to 50.0 V applied at the current output. Pulse output Displays the value currently measured for...
Operation Proline Prowirl F 200 PROFIBUS PA Parameter overview with brief description Parameter Prerequsite Description Selection / User Factory setting entry Control Totalizer In the Assign process variable Control totalizer value. • Totalize Totalize parameter one of the following • Reset + hold options is selected: •...
Page 109
Proline Prowirl F 200 PROFIBUS PA Operation If the length of the logging interval or the assignment of the process variables to the channels is changed, the content of the data logging is deleted. Navigation "Diagnostics" menu → Data logging "Data logging"...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting 12.1 General troubleshooting For local display Problem Possible causes Remedy Local display dark and no output Supply voltage does not match that Apply the correct supply voltage signals specified on the nameplate.
Page 111
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting For access Problem Possible causes Remedy No write access to parameters Hardware write protection enabled Set the write protection switch on the main electronics module to the OFF position (→ 102).
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA 12.2 Diagnostic information on local display 12.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
Page 113
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. A0013961 • For local display with touch control: the background lighting changes to red.
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA 12.2.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA 1. Call up the desired parameter. 2. On the right in the working area, mouse over the parameter. A tool tip with remedy information for the diagnostic event appears. 12.4 Adapting the diagnostic information 12.4.1...
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting Mnemonic Octet Maintenance required Maintenance alarm Maintenance demanded Function check Invalid process conditions 12.4.2 Adapting the diagnostic behavior Each item of diagnostic information is assigned a specific diagnostic behavior at the factory.
Page 118
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Coding Measuring value (hex) Byte 5 Quality Quality Substatus Limits A0021271-EN 30 Structure of the coding byte The content of the coding byte depends on the configured failsafe mode in the particular function block.
Page 119
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting Diagnostic information pertaining to the electronics (diagnostic no.: 200 to 399) Measured value status (fixed assignment) Diagnostic behavior Device diagnostics Quality Coding Category (configurable) (fixed assignment) Quality Substatus (hex) (NE107) Alarm...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA 12.5 Overview of diagnostic information The amount of diagnostic information and the number of measured variables affected increase if the measuring device has one or more application packages. 12.5.1 Diagnostics for the sensor Diagnostic no.
Page 121
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting Diagnostic no. 062 Message: F062 Sensor connection Measured value status Measured variables concerned Quality: • Calculated saturated steam pressure • Energy flow Quality Substatus: Maintenance alarm, more diagnosis • Flow velocity available •...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA • Low flow cut off Coding (hex): 0x24 • Mass flow Category (NE107): F (Failure) • Total mass flow • Status value pulse/freq./switch output • Corrected volume flow Diagnostic behavior: Alarm •...
Page 123
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting • Volume flow • Check electronic modules. • Change I/O or main electronic module. Diagnostic no. 261 Message: F261 electronic modules Measured value status Measured variables concerned Quality: • Calculated saturated steam pressure •...
Page 124
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Quality: • Calculated saturated steam pressure • Energy flow Quality Substatus: Maintenance alarm, more diagnosis • Flow velocity available • Heat flow difference • Low flow cut off Coding (hex): 0x24 •...
Page 125
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting • Steam quality Remedial measures • Temperature Change I/O module. • Volume flow Diagnostic no. 276 Message: F276 I/O module failure Measured value status Measured variables concerned Quality: • Calculated saturated steam pressure •...
Page 126
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Quality: • Calculated saturated steam pressure • Energy flow Quality Substatus: Maintenance alarm, more diagnosis • Flow velocity available • Heat flow difference • Low flow cut off Coding (hex): 0x24 •...
Page 127
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting • Volume flow Maintenance required! • Do not reset device. • Contact service. Diagnostic no. Message: *350 Pre-amplifier defective Measured value status Measured variables concerned Quality: • Calculated saturated steam pressure •...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Measured value status Measured variables concerned Quality: • Calculated saturated steam pressure • Energy flow Quality Substatus: Maintenance alarm, more diagnosis • Flow velocity available • Heat flow difference • Low flow cut off...
Page 129
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting • Low flow cut off Category (NE107): C (Check) • Mass flow • Total mass flow Diagnostic behavior: Warning • Status value pulse/freq./switch output • Corrected volume flow Remedial measures • Steam quality Up-/download active: •...
Page 130
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Diagnostic no. 482 Message: F482 Block in OOS Measured value status Measured variables concerned Quality: – Quality Substatus: Maintenance alarm, more diagnosis available Coding (hex): 0x24 Category (NE107): F (Failure) Diagnostic behavior:...
Page 131
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting • Steam quality Remedial measures • Temperature Switch off frequency output simulation. • Volume flow Diagnostic no. 493 Message: C493 Simulation pulse output Measured value status Measured variables concerned Quality: GOOD •...
Page 132
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Category (NE107): C (Check) Diagnostic behavior: Warning Remedial measures Deactivate simulation. Diagnostic no. 538 Message: S538 Computer configuration incorrect Measured value status Measured variables concerned Quality: GOOD • Calculated saturated steam pressure •...
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting 12.5.4 Diagnostics for the process Diagnostic no. 801 Message: Supply voltage too low Measured value status Measured variables concerned Quality: UNCERTAIN • Calculated saturated steam pressure • Energy flow Quality Substatus: Process related, no maintenance •...
Page 134
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA • Total mass flow Diagnostic behavior, configurable (→ 117) • Status value pulse/freq./switch output Warning (factory setting) • Corrected volume flow • Steam quality Remedial measures • Temperature • Volume flow Reduce ambient temperature of pre-amplifier.
Page 135
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting Quality: UNCERTAIN • Calculated saturated steam pressure • Energy flow Quality Substatus: Process related, no maintenance • Flow velocity • Heat flow difference Coding (hex): 0x78 • Low flow cut off...
Page 136
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA Reduce flow velocity. Diagnostic no. 862 Message: S862 Partly filled pipe Measured value status Measured variables concerned Quality: GOOD • Calculated saturated steam pressure • Energy flow Quality Substatus: • Flow velocity •...
Page 137
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting • Total mass flow Coding (hex): 0x78 • Status value pulse/freq./switch output Category (NE107): S (Out of specification) • Corrected volume flow • Steam quality Diagnostic behavior, configurable (→ 117)
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting To call up the measures to rectify a diagnostic event: • Via local display (→ 114) • Via "FieldCare" operating tool (→ 115) Other pending diagnostic events can be displayed in the Diagnostic list submenu(→...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA 12.8 Event logbook 12.8.1 Event history / ../Eventlist I1091 Config. change I1157 Mem.err. ev.list 0d01h19m10s F311 Electr. failure A0014008-EN 32 Illustrated using the example of the local display To call up the measures to rectify a diagnostic event: •...
Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting Info number Info name I1187 Settings downloaded with display I1188 Display data cleared I1189 Backup compared I1209 Density adjustment ok I1221 Zero point adjust failure I1222 Zero point adjustment ok I1256...
Diagnostics and troubleshooting Proline Prowirl F 200 PROFIBUS PA 12.10 Device information The Device information submenu contains all the parameters that display different information for identifying the device. Navigation "Diagnostics" menu → Device information Device information → Device tag Serial number...
"Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download • Specify the following details: –...
Endress+Hauser housing seals. 13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. For a list of some of the measuring and test equipment, refer to the "Accessories"...
• Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices by Endress+Hauser Service or at the factory only. Notes for repair and conversion For repair and modification of a measuring device, observe the following notes: •...
• Can be read out via the Serial number parameter in the Device information submenu (→ 142). 14.3 Endress+Hauser services Contact your Endress+Hauser Sales Center for information on services and spare parts. 14.4 Return The measuring device must be returned if repairs or a factory calibration are required, or if the wrong measuring device has been ordered or delivered.
Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
Proline Prowirl F 200 PROFIBUS PA Technical data Technical data 16.1 Application Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media. To ensure that the device remains in proper operating condition for its service life, use the measuring device only for media against which the process-wetted materials are adequately resistant.
Page 150
Technical data Proline Prowirl F 200 PROFIBUS PA Measuring range The measuring range depends on the fluid and nominal diameter. Lower range value Depends on the density and the Reynolds number (Re = 5 000, Re = 20 000). The...
Proline Prowirl F 200 PROFIBUS PA Technical data External pressure compensation is recommended to calculate the following measured variables: • Mass flow • Corrected volume flow • Energy flow 16.4 Output Output signal Pulse/frequency/switch output Function Can be set to pulse, frequency or switch output...
Page 152
Technical data Proline Prowirl F 200 PROFIBUS PA Number of switching Unlimited cycles Assignable functions • Off • On • Diagnostic behavior • Limit value – Volume flow – Corrected volume flow – Mass flow – Flow velocity – Temperature –...
Page 153
Proline Prowirl F 200 PROFIBUS PA Technical data Status signal as per NAMUR recommendation NE 107 Operating tool • Via digital communication: PROFIBUS PA • Via service interface Plain text display With information on cause and remedial measures Ex connection data...
Page 154
Technical data Proline Prowirl F 200 PROFIBUS PA Type of protection Ex ic Order code for "Output" Output type Intrinsically safe values Option G PROFIBUS PA STANDARD FISCO = 32 V = 17.5 V = 300 mA = n.a. = n.a.
Proline Prowirl F 200 PROFIBUS PA Technical data Input values Analog output (from automation system to External pressure, gage pressure, density, temperature or second temperature measuring device) (for delta heat measurement) Digital output 1 to 3 (fixed assignment) • Digital output 1: switch positive zero return on/off •...
Page 156
Technical data Proline Prowirl F 200 PROFIBUS PA Power consumption Transmitter Order code for "Output" Maximum power consumption Option G: PROFIBUS PA, pulse/frequency/ • Operation with output 1: 512 mW switch output • Operation with output 1 and 2: 2 512 mW...
Proline Prowirl F 200 PROFIBUS PA Technical data Depending on the temperature class, restrictions apply to the ambient temperature for device versions with overvoltage protection (→ 159)(→ 21) 16.6 Performance characteristics Reference operating • Error limits following ISO/DIN 11631 conditions •...
Page 158
• Re between 5 000 to 20 000: < 0.85 % o.f.s. Mass flow (user-defined liquids) • To specify the system accuracy, Endress+Hauser requires information about the type of liquid and its operating temperature or information in table form about the dependency between the liquid density and the temperature.
Proline Prowirl F 200 PROFIBUS PA Technical data Flange connection: • DN 15 (½"): ±20 % of the internal diameter • DN 25 (1"): ±15 % of the internal diameter • DN 40 (1½"): ±12 % of the internal diameter •...
Technical data Proline Prowirl F 200 PROFIBUS PA Temperature tables Observe the interdependencies between the permitted ambient and fluid temperatures when operating the device in hazardous areas. For detailed information about the temperature tables, refer to the Technical Information for the device on the CD-ROM provided...
Proline Prowirl F 200 PROFIBUS PA Technical data DSC-Sensor Order code for "Sensor Option": • Option CD "Harsh environment, DSC sensor components Alloy C22": –200 to +400 °C (–328 to +752 °F), DSC sensor Alloy C22 • Option CE "Harsh process, wetted parts Alloy C22, (Option CD included)": –40 to +260 °C (–40 to +500 °F), sensor and DSC sensor Alloy C22...
Page 162
Technical data Proline Prowirl F 200 PROFIBUS PA Weight [kg] [mm] Order characteristic "housing," option C Order characteristic "housing," option B Alu coated 316L 111.1 113.8 158.1 160.8 For high-temperature/low-temperature version: values + 0.2 kg Weight in US units All values (weight) refer to devices with ASME B16.5, Class 300/Sch. 40 flanges. Weight information in [lbs].
Page 163
Proline Prowirl F 200 PROFIBUS PA Technical data Weight in SI units All values (weight) refer to devices with EN (DIN), PN 40 flanges. Weight information in [kg]. Weight [kg] [mm] Order characteristic "housing," option C Order characteristic "housing," option B...
Page 164
Technical data Proline Prowirl F 200 PROFIBUS PA Accessories Flow conditioner Weight in SI units Pressure rating Weight [mm] [kg] PN 10 to 40 0.04 PN 10 to 40 PN 10 to 40 PN 10 to 40 PN 10 to 40...
Page 165
Proline Prowirl F 200 PROFIBUS PA Technical data Pressure rating Weight [mm] [kg] 0.06 1.80 15.8 19.1 26.5 Weight in US units Pressure rating Weight [in] [lbs] ½ Class 150 0.07 Class 300 0.09 Class 150 Class 300 1½ Class 150...
Page 166
Technical data Proline Prowirl F 200 PROFIBUS PA Materials Transmitter housing Compact version • Order code for "Housing", option C "Compact, aluminum coated": Coated aluminum AlSi10Mg • Order code for "Housing", option B "Compact, stainless": For maximum corrosion resistance: stainless steel 1.4404 (316L) Remote version •...
Page 167
Proline Prowirl F 200 PROFIBUS PA Technical data Cable entry/cable gland Type of protection Material Adapter for cable entry with For non-Ex and Ex Nickel-plated brass internal thread NPT ½" (except for CSA Ex d/XP) Thread NPT ½" For non-Ex and Ex...
Technical data Proline Prowirl F 200 PROFIBUS PA Seals • Graphite (standard) Pressure rating PN 10 to 40, Class 150 to 300, JIS 10 to 20K: Sigraflex Foil Z (BAM- certified for oxygen applications) • FPM (Viton) • Kalrez 6375 •...
EC Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
The application packages can be ordered from Endress+Hauser either directly with the device or subsequently. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
Overview of accessories available for order (→ 147) 16.15 Documentation The following document types are available: • On the CD-ROM supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download Standard documentation Communication...
Appendix Proline Prowirl F 200 PROFIBUS PA Appendix 17.1 Overview of the operating menu The following table provides an overview of the entire operating menu structure with menus and parameters. The direct access code to the parameter is given in brackets. The page reference indicates where a description of the parameter can be found in the manual.
Page 173
Proline Prowirl F 200 PROFIBUS PA Appendix Select gas type (→ 73) (→ 74) Select liquid type (→ 74) Select steam type Enthalpy calculation (→ 74) Density calculation (→ 74) Enthalpy type (→ 74) System units →...
Page 174
Appendix Proline Prowirl F 200 PROFIBUS PA Display → (→ 75) Format display (→ 77) Value 1 display (→ 77) 0% bargraph value 1 (→ 77) 100% bargraph value 1 (→ 77) Value 2 display (→...
Page 175
Proline Prowirl F 200 PROFIBUS PA Appendix Dynamic viscosity (→ 81) Gas composition → (→ 81) Gas type (→ 83) Gas mixture (→ 83) Mol% Ar (→ 83) Mol% C2H3Cl (→ 83) Mol% C2H4 (→...
Page 177
Proline Prowirl F 200 PROFIBUS PA Appendix (→ 92) Switch output function Assign diagnostic (→ 92) behavior (→ 93) Assign limit Assign flow direction (→ 93) check Assign status (→ 94) (→ 71) Mass flow unit Mass unit (→...
Page 178
Appendix Proline Prowirl F 200 PROFIBUS PA Measuring value at (→ 90) minimum frequency Failure mode (→ 91) Failure frequency (→ 91) Switch-on value (→ 93) Switch-off value (→ 94) Switch-off value (→ 94) Switch-on value (→...
Appendix Proline Prowirl F 200 PROFIBUS PA 17.1.4 "Diagnostics" menu Diagnostics → (→ 110) Actual diagnostics (→ 139) Previous diagnostics (→ 139) Operating time from (→ 139) restart Operating time (→ 140) Diagnostic list →...
Page 181
Proline Prowirl F 200 PROFIBUS PA Appendix Heat flow difference (→ 105) (→ 105) Reynolds number (→ 105) Density Pressure (→ 105) Compressibility factor (→ 105) Totalizer 1 to 3 → (→ 105) Assign process variable (→...
Page 182
Appendix Proline Prowirl F 200 PROFIBUS PA AM/PM Minute Verification mode External device information Start verification Status Measured values Output values Overall result Verification results → Date/time Verification ID Operating time Overall result Sensor Pre-amplifier module Main electronic module I/O module Simulation →...
Page 183
Proline Prowirl F 200 PROFIBUS PA Appendix Simulation diagnostic (→ 101) event Order characteristic "Application package", option EA "Extended HistoROM", see Technical Information for device, "Application packages" section Order code for "Application package", option EB "Heartbeat Verification", see the Special Documentation for the device...
Page 192
Appendix Proline Prowirl F 200 PROFIBUS PA Calibration factor (7604) Meter body properties (7658) Year (2846) Month (2845) Day (2842) Hour (2843) AM/PM (2813) Minute (2844) "Output" submenu → Output Pulse-Frequency- → (→ 86) Switch output Operating mode (0469) (→...
Page 193
Proline Prowirl F 200 PROFIBUS PA Appendix Output frequency (0471) (→ 107) (→ 92) Switch output function (0481) Assign diagnostic (→ 92) behavior (0482) Assign limit (0483) (→ 93) Switch-on value (0466) (→ 93) Switch-off value (0464) (→...
Page 194
Appendix Proline Prowirl F 200 PROFIBUS PA Device tag (1496) (→ 70) Static revision (1495) Strategy (1494) Alert key (1473) Target mode (1497) Mode block actual (1472) Mode block permitted (1493) Mode block normal (1492) Alarm summary (1474) Software revision (1478)
Page 195
Proline Prowirl F 200 PROFIBUS PA Appendix "Analog inputs" submenu Analog inputs → Analog input 1 to 4 → (→ 75) Channel (1561) (→ 75) PV filter time (1524) (→ 75) Fail safe type (1525) (→ 75) (→...
Page 196
Appendix Proline Prowirl F 200 PROFIBUS PA Out status (1669) Out status (1645) "Discrete outputs" submenu → Discrete outputs Discrete output 1 to 3 → Set point value (1715) Set point status (1714) Invert (1692) Fail safe time (1697) Fail safe type (1696)
Page 197
Proline Prowirl F 200 PROFIBUS PA Appendix "Diagnostics" submenu Diagnostics → (→ 110) Actual diagnostics (→ 139) (0691) (→ 139) Previous diagnostics (0690) Operating time from (→ 139) restart (0653) Operating time (0652) (→ 140) →...
Page 198
Appendix Proline Prowirl F 200 PROFIBUS PA Minimum value (0689) Maximum value (0663) Average value (0698) IO module temperature → Minimum value (0688) Maximum value (0665) Average value (0697) Pre-amplifier → temperature Minimum value (7724) Maximum value (7723) Medium temperature →...
Page 199
Proline Prowirl F 200 PROFIBUS PA Appendix Start verification (12127) Status Measured values (12102) Output values Overall result (12149) Verification results → Date/time (12142) Verification ID (12141) Operating time (12126) Overall result (12149) Sensor (12152) Pre-amplifier module (12151) Main electronic module...
Page 200
Appendix Proline Prowirl F 200 PROFIBUS PA Simulation diagnostic (→ 101) event (0737) Order characteristic "Application package", option EA "Extended HistoROM", see Technical Information for device, "Application packages" section Order code for "Application package", option EB "Heartbeat Verification", see the Special Documentation for the device...
Proline Prowirl F 200 PROFIBUS PA Index Index Degree of protection ..... 44, 160 Design Access authorization to parameters Measuring device ......11 Read access .
Page 202
Index Proline Prowirl F 200 PROFIBUS PA Operating tools Information on the document ....6 Via PROFIBUS PA network ....61 Inlet runs .
Page 203
Proline Prowirl F 200 PROFIBUS PA Index Media ........9 Configuration backup display (submenu) .
Page 204
Index Proline Prowirl F 200 PROFIBUS PA Response time ......159 Display ....... . . 96 Returning devices .
Page 205
Proline Prowirl F 200 PROFIBUS PA Index Troubleshooting General ....... 110 Turning the display module .
Need help?
Do you have a question about the Proline Prowirl F 200 and is the answer not in the manual?
Questions and answers