Instructions for use

Security devices

Emergency stop:
The Poweo
®
patient lifters are equipped with an emergency stop button that
will turn off the lifter when pressed.
The emergency stop button is located on the top of the battery on the
Poweo
®
140 -180 Quick Up. Once pressed It must be turned clockwise to
release it.
Emergency lowering devices:
In case the remote control does not operates correctly, an emergency electric
lowering is still possible by activating the hidden button located behind the
yellow label of the base.
Perforate the center of the yellow sticker and press the hidden button with a
small screwdriver and the patient lifter will perform an emergency lowering.
Maintain the pressure on the button to obtain the desired height.
In case of a complete failure of the electronics, a mechanical emergency
lowering is available.
Activate the RED lever/trigger on the actuator by pulling it upward, and
maintain the position until the actuator reaches the desired height.
Lowering speed will depend on the charge and the pressure on the trigger. In case of
use, make sure that a chair or a bed can receive the patient when he is being lowered.
According to the ISO EN 10535 standard, the Poweo
weighing more than its maximum safety load capacity.