Manual URYXXON
Activating the entry "Strip LOT" enables the download of strip LOT data at each
start of the instrument. The strip LOT data entered before the measurement are
checked against these downloaded data (Chapter 10.6 "How to change the QC
measurement information").
Activating the entry "Control LOT" enables the download of control LOT data
at each start of the instrument. The strip LOT data entered before the mea-
surement are checked against these downloaded data (Chapter 10.6 "How to
change the QC measurement information").
Activating the entry "Comments" enables the download of a list of predefined
comments from the remote system. These comments will be displayed in a list
when entering a comment after the measurement when choosing the comment
option (see Chapter 5.4 "Enter a comment to a measurement"). The instrument
can download up to seven predefined comments with a length of up to 30 char-
acters.
Activating the entry "System message" enables the download of a system mes-
sage from the remote system. This message will be displayed after the self-test
during instrument start. The user needs to confirm this message. The message
is limited to 100 characters.
Activating the entry "User message" enables the download of a user specific
message from the remote system. This message will be displayed for the spe-
cific user after login with the respective user name. The user needs to confirm
this message. The message is limited to 100 characters.

10.12. QC lockout

In bidirectional mode, the instrument can be locked from remote. In this case the
instrument displays the message "locked from remote" after the instrument start.
In this status, it is still possible to perform an emergency measurement and a
QC measurement depending on the user roles. When performing an emergency
measurement, the comment "QC override" is automatically attached to the mea-
surement result.
NOTE:
The instrument can only be unlocked via the remote system again.