How To Change The Qc Measurement Information; How To Set The Qc Evaluation; How To Change The Measurement Information; Instrument Setup - Macherey-Nagel URYXXON Relax User Manual

Medi-test
Hide thumbs Also See for URYXXON Relax:
Table of Contents

Advertisement

Manual URYXXON
instrument again.The device will be automatically unlocked after performing a
positive and a negative control measurement (see Chapter 8.2 "How to perform
a QC measurement").
When QC-evaluation is set to "Off" (see Chapter 9.12 "How to set the QC evalu-
ation"), there is no need to evaluate the QC measurement.

10.6. How to change the QC measurement information

In bidirectional mode the entered information for the LOT number and expiration
date of control solution and test strips are checked against the remote system,
in case this option is set in the remote system settings (see Chapter 10.11 "How
to change the remote settings"). When validation of the LOT number or expira-
tion date fails, using this LOT is not allowed. The user needs to use a valid LOT
number or use the emergency command to bypass the LOT and expiry check. In
case the LOT check is not supported by the remote system the LOT entry will fail
and the user needs to use the emergency command or deactivate this setting.

10.7. How to set the QC evaluation

®
The URYXXON
Relax allows for the evaluation of quality control measurements
as "Passed" and "Failed". Select the list entry "QC-Evaluation" in the menu "QC
setup" and press the round button "QC user" to activate the QC evaluation. The
default setting is "Off".
ATTENTION:
In bidirectional mode the POC coordinator should evaluate, which measures
for QC evaluation need to be taken. A QC evaluation can happen by the QC-
user or also via the remote system.

10.8. How to change the measurement information

For general information on the measurement information settings please refer to
Chapter 9.13 "How to change the measurement information".
In bidirectional mode the entered information for the LOT number and expiration
date of control solution and test strips are checked against the remote system,
in case this option is set in the remote system settings (see Chapter 10.11 "How
to change the remote settings"). When validation of the LOT number or expira-
tion date fails, using this LOT is not allowed. User needs to use a valid LOT
number or use the emergency command to bypass the LOT and expiry check.

10.9. Instrument setup

Within this menu entry the setup for the instrument can be changed.
For general information on the instrument setup please refer to Chapter 9.14
"Instrument setup".
48
®
Relax
Manual URYXXON
®
Relax EN, V 5.00C / 06.19