To release the instrument for patient testing, please perform a positive and a
negative QC measurement by pressing
evaluated as "Passed".
Alternatively the QC lockout can be bypassed by pressing
"QC override" will be displayed shortly. The device allows to run a patient sample
now, but the comment "QC override" will be added automatically to the measure-
ment to assign that this measurement has been performed during QC lockout
status.
NOTE:
To release the instrument for patient testing, please perform a positive and
a negative QC measurement. Both measurements need to be evaluated as
"Passed".

9.11. How to change the QC measurement information

Besides the ID for negative and positive quality control measurement, the user
can activate the option to obligatory assign the LOT number and expiration date
of the used strips and control solution. To activate the respective information
select the list entry "QC information" in the menu "QC setup" and use the arrow
keys and
to change the setting.
Activating the list entries "LOT" and/or "Control LOT" requires the input for the
LOT and expiry information of the used strips and control solution prior to the
QC measurement. The information can be entered via the alphanumerical pad
on the display or an external keyboard.
40
®
Relax
URYXXON STICK 10
QC period lapsed!
Please run QC.
(P)
Display 30: QC lockout
. Both measurements need to be
Manual URYXXON
. The message
®
Relax EN, V 5.00C / 06.19