entered user name and password are validated against a user list on the remote
system. After successful validation, the user can operate the instrument accord-
ing to the activated user roles.
NOTE:
In bidirectional mode the user management is only possible, when a user list
is active on the remote system.

10.3. How to print a user list

To print a list of users select the list entry "Print users" in the menu "User man-
agement". The instruments prints a list of the first 100 users stored on the re-
mote system with their respective roles and the user status (Active or Inactive).
The order of the user list is based on the order in the remote system.

10.4. QC-setup

For general information on the QC-setup please refer to Chapter 9.9 "QC-setup".

10.5. How to set a QC reminder

When activated the device will display the advisory message "QC measure-
ment required" each time the reminder is due (Display 32). The instrument will
be locked for patient testing. An emergency measurement is still possible, if this
setting is activated.
To release the instrument for patient testing, please perform a positive and a
negative QC measurement by pressing
When QC-evaluation is set to "QC user" (see Chapter 9.12 "How to set the
QC evaluation"), the results need to be evaluated by the user as being passed
or failed. Both measurements need to be evaluated as "Passed" to unlock the
Manual URYXXON
Manual URYXXON
URYXXON STICK 10
QC period lapsed!
Please run QC.
(P)
Display 32: QC lockout
®