8. Quality control testing
Quality control measurements should be performed regularly with check solu-
tions in order to ensure the correct functioning of the combination of equipment
and test strips.
Every facility has to implement its own QC policy.
Upon pressing
? on the start menu the "Check mode" screen will appear
(Display 21 + Display 22).
MENU
Memory
?
Check mode
Settings
Display 21: Main menu
8.1. How to review old QC measurements
The equipment saves the results of the last 20 QC measurements in a sepa-
rate memory. They may be displayed by pressing
printed for documentation purposes.
32
®
Relax
Manual URYXXON
CHECK MODE
Memory
Measurement
Display 22: Check mode
(Display 22) and may be
®