10.10. Other settings

10.10.1. How to activate the user accounts and user password

Selecting "User accounts" in the menu "Other settings" will activate the user
mode of the instrument. After the instrument start, a user login is required. If
the entry "User password" has been activated additionally, there will also be a
password required for the user login at instrument start.
ATTENTION:
When activating the "User accounts" make sure that the device is connected
to a remote system, which is prepared to run the user authentication process
with the URYXXON
user and only offers the possibility of emergency measurements!

10.11. How to change the remote settings

The menu "Remote setting" is only available in bidirectional interface mode (see
Chapter 9.22 "How to change the interface settings"). Selecting the menu opens
a selection list to set the different remote settings with a check box (Display 33).
Activating the entry "PPID" allows the request of patient information from the
remote system. After entering the patient ID and confirmation, the instruments
displays the patient information name, surname, date of birth (DoB) and gender.
By confirming the information, the user positively identifies that the ID matches
the patient.
Activating the entry "User" allows for a user validation against the remote sys-
tem (see Chapter 9.5 "User management").
Manual URYXXON
®
Relax. Otherwise, the device cannot be accessed by a
URYXXON STICK 10
REMOTE COMM.
PPID
Users
Strip LOT
Control LOT
Comments
System message
User message
Display 33: Set remote comm.
®