9.8. How to print a user list

To print a user list select the list entry "Print users" in the menu "User manage-
ment". A list of all stored users with their respective roles and the user status
(active or inactive) is printed. The order of the user list is based on the date when
the user has been created.

9.9. QC-setup

The menu QC-setup allows to change the settings for QC measurements. Be-
side setting a QC reminder it is possible to activate an evaluation for the QC
measurements. Furthermore a protocol for the required QC measurement infor-
mation can be activated.

9.10. How to set a QC reminder

To activate a popup message as a reminder for quality control measurements,
please choose the list entry "QC reminder" in the menu "QC setup". To activate
the QC reminder press the round button "On" (Display 29).
The reminder is disabled by default. Press the Icons "Count" and "Interval" to
choose the interval for the reminder. By combination of the two entries intervals
starting from 1 hour to several years are possible. Pressing
entries.
When activated the device will display the advisory message "QC measurement
required" each time the reminder is due (Display 30). The instrument will be
locked for patient testing. Only emergency measurements are possible now, if
this setting has been activated.
Manual URYXXON
Manual URYXXON
QC REMINDER
QC lock
On
Off
Count
1
Interval
Hours
Display 29: Set QC reminder
®
Relax EN, V 5.00C / 06.19
®