Rh Set Point Is Not Achieved; Bake Temperature Set Point Is Not Achieved - ECD SmartDRY SD Series Setup And Operation Manual

Dry cabinets
Hide thumbs Also See for SmartDRY SD Series:
Table of Contents

Advertisement

11.2 RH SET POINT IS NOT ACHIEVED:

RH Set point is not set to the value you want:
o
Confirm RH set point value.
The door is not closed or sealed well. Although the door alarm will sound if a door is left open, leaking
ambient air can be caused by:
o
Check for broken or distorted door seals. Distorted door seal due to shipping will return to shape
in time once returned to the normal position. Damaged seal will need to be replaced; contact ECD
for help with this.
o
The Cabinet is on an uneven surface and one or more doors is not able to seal.
o
Make sure nothing is defeating the door closed switches.
Dryer is not full regenerated yet:
o
Most likely the Dryer has not fully regenerated. This takes about 2 hours and if the ECD
SmartDRY™ Cabinet has been off for a long period during shipping or while out of service, it will
require both Dryer Modules inside the Dryer be regenerated before they can be fully effective.
Give it at least 4 hours to complete the Regeneration process.
The Dryer had been damaged or has a component that has failed:
o
Replace the Dryer following the procedure outlined in the Section
The RH/Temperature Probe is damaged:
o
Replace the probe following the procedure outlined in the Maintenance section: RH/Temp Probe.

11.3 BAKE TEMPERATURE SET POINT IS NOT ACHIEVED:

Bake Temperature set point is not set to the value you want:
o
Confirm Bake Temperature set point value.
Bake Time has completed or the Bake time was not set to the expected value:
o
Confirm Bake time value.
o
The product in the cabinet is "High-Mass" and may take longer to process.
You may not have purchased the ECD SmartDRY™ 3D-10SB cabinet with SmartBAKE™. Normally the
Controller will not allow you to program the SmartBAKE™ cycle unless you purchased this cabinet.
However, if SmartBAKE™ is in your display menu and you did not purchase it, contact ECD and we can
correct that error in configuration.
The Bake heater may be damaged.
The RH/Temperature Probe is damaged:
Replace the probe following the procedure outlined in the Maintenance section: RH/Temp Probe.
o
SD Series Dry Cabinets Setup and Operation Manual | A62-3337-00 Rev: 1.0.0