11.2 RH SET POINT IS NOT ACHIEVED:

RH Set point is not set to the value you want:
Confirm RH set point value.
o
The door is not closed or sealed well. Although the door alarm will sound if a door is left open, leaking
ambient air can be caused by:
Check for broken or distorted door seals. Distorted door seal due to shipping will return to shape
o
in time once returned to the normal position. Damaged seal will been to be replaced; contact ECD
for help with this.
The Cabinet is on an uneven surface and one or more doors is not able to seal.
o
Make sure nothing is defeating the door closed switches
o
Dryer is not full regenerated yet:
Most likely the Dryer has not fully regenerated. This takes about 2 hours and if the SmartDRY™
o
Cabinet has been off for a long period during shipping or while out of service, it will require both
Dryer Modules inside the Dryer be regenerated before they can be fully effective. Give it at least 4
hours to complete the Regeneration process.
The Dryer had been damaged or has a component that has failed:
Replace the Dryer following the procedure outlined in the Section
o
The RH/Temperature Probe is damaged:
Replace the probe following the procedure outlined in the Maintenance section: RH/Temp Probe
o