Auto Tests

The defibrillator performs automatic self-tests daily and monthly at 0300 (3:00 a.m.) if not in use.
During the automatic self-test, the defibrillator turns itself on (ON/OFF LED illuminates) briefly
and completes the following tasks:
• Performs a self-test
• Stores the self-test results in the Test Log
• Turns itself off
If the defibrillator detects a problem during an auto test that requires service, it displays the
service symbol. If the service symbol is visible, you should attempt to use the defibrillator, if
needed, for a cardiac emergency. However, you should contact authorized service personnel to
correct the problem as soon as possible. The service symbol will remain visible until the
problem is corrected.
The automatic self-test is not performed if the defibrillator is already turned on at 0300 or if the
battery is not installed. If the defibrillator is turned on while a self-test is in progress, the test is
halted; the defibrillator will turn on normally.

INSPECTION

Routinely inspect all devices, accessories, and cables by following the instructions in Table 5-2.
LIFEPAK 1000 Defibrillator Inspection
Table 5-2
Instruction
Inspect For
Examine the
Foreign substances.
defibrillator case,
connector, battery well,
Damage or cracks.
battery pins, and
accessories.
Battery pins bent or
discolored.
Expired batteries or
defibrillation electrodes.
Observe readiness
OK
display
Battery status indicator
shows one bar or less.
Service symbol displayed.
Examine accessory
Foreign substances.
cables.
Inspect for cracks, damage,
extreme wear, broken or bent
connectors and pins.
Confirm that connectors
engage securely.
LIFEPAK 1000 Defibrillator Operating Instructions
Copyright ©2021 Stryker
Caring for the LIFEPAK 1000 Defibrillator
Recommended Corrective Action
Clean the device as described in
Table
Contact authorized service personnel
to troubleshoot.
Contact authorized service personnel.
Replace.
symbol.
None needed.
Replace battery immediately.
Contact authorized service personnel.
Clean the cables as described in
Table
Replace damaged or broken parts.
Replace damaged or broken parts.
5-3.
5-3.
5-3