Troubleshooting Tips for ECG Monitoring Mode

If problems occur while monitoring the ECG, check this list of observations for troubleshooting
assistance.
Troubleshooting Tips for ECG Monitoring Mode
Table 3-2
Observation
Screen blank and ON
LED lit.
CONNECT ECG LEADS
message appears
3-10
Possible Cause
Screen not functioning
properly.
One or more ECG electrodes
are disconnected.
Poor electrode-to-skin
contact.
Broken ECG cable lead wire.
What to Do
• Contact authorized service
personnel.
• AED and therapy functions may
still operate. If needed for
therapy, continue to use device
to treat patient.
• Confirm ECG electrode
connections.
• Reposition cable and/or lead
wires to prevent electrodes from
pulling away from patient.
• Clean, shave, and dry the
patient's skin as recommended
on
page
3-3.
• Replace electrodes.
• Change cable.
• Check ECG cable continuity. If
lead wire is broken, replace ECG
cable.
LIFEPAK 1000 Defibrillator Operating Instructions