Inspection checklist
Refer to the Operating
Instructions for the
User's Checklist
• Inspect the defibrillator for
signs of damage, cracks or
foreign substances
• It is not necessary to switch
on the defibrillator for the
inspection
• Observe the Readiness Display
on the top right and perform
the following checks:
1. The OK symbol is visible
2. The battery symbol shows
two or more bars
• Confirm spare, fully charged
battery is readily available
and in-date
• Ensure there is one pack
of QUIK-COMBO pads
preconnected to the
defibrillator, and the pack of
pads is sealed and in-date
• There should be one
additional set of adult pads
• If present, check Infant/Child
Reduced Energy Defibrillation
Electrodes to make sure they
are sealed and in-date
Readiness
display
OK indicates device
is ready for use.
Battery is fully charged.
Battery is very low.
Replace immediately.
Battery is depleted.
Replace immediately.
Service symbol indicates
service is required.
LIFEPAK
1000
®
Pocket
guide
This is not the complete instructions for use. Refer to
Operating Instructions for complete device information
including directions for use, intended use and warnings.
Defibrillator

Advertisement

loading
Need help?

Need help?

Do you have a question about the LIFEPAK 1000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Stryker LIFEPAK 1000

  • Page 1 Inspection checklist Refer to the Operating Instructions for the User’s Checklist LIFEPAK 1000 Defibrillator ® • Inspect the defibrillator for Readiness signs of damage, cracks or display foreign substances Pocket guide • It is not necessary to switch on the defibrillator for the OK indicates device is ready for use.
  • Page 2 For a brief summary of indications and important safety information, visit our website at www.strykeremergencycare.com/disclosure-and-safety-information/lifepak-1000-defibrillator/ For further information, please contact your local Stryker representative or visit our website at www.strykeremergencycare.com Physio-Control, Inc., 11811 Willows Road NE, Redmond, WA 98052 U.S.A.