Ecg Monitoring (Ecg Mode) - Stryker LIFEPAK 1000 Operating Instructions Manual

Defibrillator
Hide thumbs Also See for LIFEPAK 1000:
Table of Contents

Advertisement

ECG MONITORING (ECG MODE)

WARNING!
Possible misinterpretation of ECG data.
The frequency response of the screen is intended only for basic ECG rhythm identification; it
does not provide the resolution required for pacemaker pulse visibility, accurate measurements,
such as QRS duration, and ST segment interpretation. For such purposes, use ECG monitors
with an appropriate frequency response.
Possible delay in therapy.
Do not attempt to connect a 3-wire ECG cable to a QUIK-COMBO therapy cable or any other
AED. The ECG cable is functional only with the LIFEPAK 1000 defibrillator.
The LIFEPAK 1000 defibrillator provides nondiagnostic ECG display of the patient's heart
rhythm when the ECG cable is connected and the electrodes are applied.
Note: You do not have to turn the defibrillator off before changing from therapy electrodes
to the ECG cable or vice versa.
To use ECG monitoring mode:
1 Connect the ECG cable.
Note: The ECG cable uses the same receptacle used by the therapy electrodes.
2 Apply ECG electrodes to the patient's chest as shown in
AHA Labels
RA
LA
LL
RA/R
LA/L
LL/F
Connecting the ECG Electrodes for use with ECG cable
Figure 3-2
After the ECG electrodes are connected, the defibrillator displays the patient's heart rhythm and
heart rate in a lead II configuration. Lead II is the only lead available with this cable.
While in ECG mode, the defibrillator's shock capability is disabled; however, the defibrillator
continues to evaluate the patient's ECG for a potentially shockable rhythm. Remember that the
presence of an ECG rhythm does not ensure that the patient has a pulse.
If a shockable rhythm is detected, the defibrillator prompts CONNECT THERAPY
ELECTRODES
.
1 Confirm the patient's condition: Not responsive? Not breathing? No signs of circulation?
2 Remove the ECG cable and connect the therapy electrodes to the defibrillator.
3 Apply the therapy electrodes to the patient's chest, keeping them at least 2.5 cm (one inch)
away from the ECG electrodes. If necessary, remove the ECG electrodes.
4 Follow the defibrillator's voice and screen prompts.
LIFEPAK 1000 Defibrillator Operating Instructions
Copyright ©2021 Stryker
How to Use the LIFEPAK 1000 Defibrillator
IEC Labels
Right Arm
R
Right
Left Arm
L
Left
Left Leg
F
Foot
Figure 3-2
3-9

Advertisement

Table of Contents
loading

Table of Contents