Manual Daily Test
First inspect the defibrillator for signs
of damage.
It is not necessary to switch the AED
on for the daily test!
Observe the display to the top right side of the
AED, and perform the following checks.
1.
The 'OK' symbol is visible.
2.
The battery and service indicator shows more
than one bar, and does not show empty
(no bars) or a spanner symbol.
3.
Check that there is one pack of in date
'Quik-Combo' pads preconnected to the
AED, and that the pack of pads is sealed
and in date.
4.
There should be one additional set
of adult Quik-Combo pads.
New Weekly Check
Remove battery and reinstall straight away
Physio-Control now part of Stryker
35 Great St Helens London, EC3A 6AP
Tel
0808 258 0094 /option 1
Battery and Service
Indicator
Battery is fully charged
Order Replacement Battery
Battery is 25% charged or less
Replace immediately
Battery is exhausted
Replace immediately
Spanner indicates
maintenance is required
urgently Contact MESU
www.physio-control.com
LIFEPAK
1000
®
Defibrillator
Pocket
Guide

Advertisement

loading

Summary of Contents for Stryker PHYSIO CONTROL LIFEPAK 1000

  • Page 1 Quik-Combo pads. New Weekly Check • Remove battery and reinstall straight away Spanner indicates maintenance is required urgently Contact MESU Physio-Control now part of Stryker 35 Great St Helens London, EC3A 6AP 0808 258 0094 /option 1 www.physio-control.com...
  • Page 2 LIFEPAK 1000 Using a LIFEPAK 1000 at a glance for AED Defibrillation Follow these 3 steps: Pre-connect pads here Battery and Service indicator Speaker Apply QUIK-COMBO REDI-PAK ® ™ electrodes to patients chest ‘On’ button GREEN ON Press the button - ‘Shock’...