Emerson Rosemount 3051P Reference Manual page 82

In-line pressure transmitter with hart protocol
Hide thumbs Also See for Rosemount 3051P:
Table of Contents

Advertisement

Troubleshooting
April 2020
Incompatible electronics and sensor
LCD display
screen
Local operator
interface (LOI)
screen
The pressure sensor is incompatible with the attached electronics.
Recommended action
Replace the pressure transmitter.
6.4.2
Status: Maintenance - Fix Soon
No temperature updates
LCD display
screen
Local operator
interface (LOI)
screen
There are no temperature updates from the sensor to the electronics.
Recommended actions
Pressure out of limits
LCD display
screen
Local operator
interface (LOI)
screen
The pressure is either above or below the sensor limits.
Recommended actions
Sensor temperature beyond limits
LOI display
screen
82
XMTR MSMTCH
XMTR MSMTCH
NO T UPDATE
NO TEMP UPDATE
1. Ensure the sensor cable connection to the electronics is tight.
2. Replace the pressure transmitter.
NO T UPDATE
PRES OUT LIMITS
1. Check the transmitter pressure connection to ensure it is not plugged and the
isolating diaphragms are not damaged.
2. Replace the pressure transmitter.
TEMP LIMITS
Reference Manual
00809-0700-4001
Emerson.com/Rosemount

Advertisement

Table of Contents
loading

Table of Contents