Performance Tests

Test method

1 Enter the recorded measurement uncertainty of the E4419B or E4417A 1 mW
2 Using the E4419B or E4417A power meter and the 8482A sensor, measure the
3 Using the E4419B or E4417A power meter and the 8482A sensor, measure the
4 Using all of these values, the N7832A software will calculate the power
– The 1 mW reference of the E4419B or E4417A power meter must be precisely
NOTE
– Anyone who has a basic understanding of metrology should be able to
24
power reference
1mW power reference of the E4419B or E4417A
1 mW power reference of the DUT
reference level of the DUT
calibrated at a standards accredited lab, and the uncertainty of this
measurement known.
perform this test manually; it is simply the transfer of known power level with
a known calibration uncertainty to the DUT.
Keysight N1911A/1912A Service Guide