5
Repair Guide

Main board vs. rear panel assembly

Instructions
– Due to a difference in the connector positions
for main board revision 102 and revision 103,
there are TWO different rear panels
– Revision 102 main boards are not available as
spares — all spare main boards will be revision
103 (or newer)
– When replacing a revision 102 main board, take
note that the rear panel will need to be replaced
Figure 5-21
and
Figure 5-22
the main board revision markings
110
Visual
Figure 5-21
show where to find
Figure 5-22
Main board revision 102
Main board revision 103