When switching the range calibrator to STANDBY, allow enough time for the
NOTE
range calibrator to settle to its zero value before attempting to zero the Keysight
N1911A/1912A P-Series power meter. This settling would appear on the
Keysight N1911A/1912A P-Series power meter display as downward drift. When
the drift has reached minimum, (typically less than 60 seconds), the range
calibrator is settled.
3 Turn on the voltmeter and allow it to warm up for 4 hours.
4 Turn on the DC source and allow it to warm up. Set the output voltage to 0 V.
5 Turn on the DUT and allow it to warm up for 30 minutes.
6 Ensure the range calibrator is not plugged-in or powered on.
7 Connect the range calibrator to the DUT using the power meter interconnect
8 Connect the DC source to the range calibrator external voltage input and the
9 Configure the DUT as shown below.
Parameter
Filter/Averaging
Filter Mode/Measurement Average
Filter Length/Average Number
Resolution
All other settings use the default setup settings. The *RST command sets the
NOTE
default setup.
10 Configure the voltmeter using the default setup settings.
Keysight N1911A/1912A Service Guide
Disconnect the power cord to stop the operation of the range calibrator as the
range calibrator will continue to operate with no power applied.
cable.
voltmeter using cables and a BNC T-joint connector at the range calibrator.
Value
On
Manual
16
4 digits
Performance Tests
2
35