Instructions
Step 10:
– Release the metal tabs holding the EMI screen
to the front panel sub-frame, and separate them
from one another
Step 11:
– Disengage the rubber tabs that attach the key
flex circuit to the keymat, and carefully lift it out
Step 12:
– Remove the EMI shielded window and the
keymat from the front panel sub-frame
Keysight N1911A/1912A Service Guide
Visual
Repair Guide
5
103