Operating The Blower; Operating The Lights; Working In Your Protector Fume Hood - Labconco Protector XL 1124 Series User Manual

Floor-mounted (walk-in) laboratory fume hood
Table of Contents

Advertisement

Original instructions

Operating the Blower

S
A
Your Protector Fume Hood utilizes a remote style blower, which can be activated
by turning the blower switch to "ON." You can validate the hood performance by
watching smoke drawn into the hood face opening.

Operating the Lights

S
A
Your Protector Fume Hood utilizes a factory-wired LED light to illuminate the
hood interior (see Appendix A). Simply turn the light switch to "ON" to operate.

Working in your Protector Fume Hood

Planning
 Thoroughly understand procedures and equipment required before
 Arrange for minimal disruptions, such as room traffic or entry into the
Start-up
Labconco follows the recommendations of the American National Standard
Institute with regards to fume hood exhaust systems. A full list of requirements
can be found in the ANSI Standard No. Z9.5 Laboratory Ventilation. If you are
unsure of the proper use of the exhaust blower for your fume hood, contact your
safety officer or industrial hygienist before making any modifications to the
blower's operational status (on/off/low, etc.). Keeping the exhaust system under
constant negative pressure is an excellent way to ensure hazardous fumes in the
fume hood do not enter the laboratory space inadvertently.
 Turn on the light.
 Slowly raise the sash and verify the fume hood has proper airflow.
 Check the baffle air slots for obstructions.
 Allow the hood to operate unobstructed for two minutes.
 Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Loading Materials and Equipment
 Only load the materials required for the procedure. Do not overload the
 Do not obstruct the front air foil (sill), or rear baffle slots.
 Large objects should not be placed close together and spaced above the
 After loading the hood, wait one minute to purge airborne contaminants
Product Service 1-800-522-7658
beginning work.
room while the hood is in use.
Wear a protective mask if appropriate.
hood.
work surface to permit airflow to sweep under the equipment.
from the work area.