Safe operation
Risk of injury and/or malfunc-
tion!
►
WARNING
activ L parallel distractor FW970R
Coupling the trial implants for anterior
access (0°)
►
Turn twist cap 10 counterclockwise until jaw
pieces 1 and 2 are in contact with each other and
the scale 4 reads less than 8.5 mm, see Fig. 14.
►
Put cranial trial implant a onto cranial jaw piece 1
in such a way that the small guide pin on the cra-
nial trial implant engages in central guide hole b of
the cranial jaw piece and audibly clicks into posi-
tion.
►
Put caudal trial implant d onto caudal jaw piece 2
in such a way that the small guide pin c on the
caudal trial implant engages in central guide hole
of the caudal jaw piece and audibly clicks into
position.
Always carry out a func-
tion check prior to using
the product.
a
b
1
2
c
d
Fig. 14 Coupling the trial implants
Legend
a Cranial trial implant
b Central guide hole
c Small guide pin
d Caudal trial implant
►
To distract, turn twist cap 10 clockwise.
►
Read the distraction height from scale 4.
Decoupling the trial implants
The trial implants can be decoupled with jaw pieces 1
and 2 opened to any width.
►
Pull at the right and left sides of trial implant a or
d, respectively while simultaneously applying
counterforce on sheath 3 of the parallel distractor,
see Fig. 14.
15