Protection against inadvertent activation
The rotational speed control button can be locked to
prevent inadvertent activation of the small drill during
a tool change.
►
To lock rotational speed control button 8: Turn
safety catch 9 to its OFF position, see Fig. 11.
Rotational speed control button 8 is locked and
small drill 6 can not be operated.
►
To unlock rotational speed control button 8: Turn
safety catch 9 to its ON position, see Fig. 11.
Rotational speed control button 8 is unlocked and
small drill 6 can be operated.
7
8
9
6
Fig. 11
Coupling tools in the tool chuck adapter
Note
The coupling of tools is described in instructions for use
TA011936 (Acculan 3Ti) and/or TA014540 (Acculan 4).
12
5.2
Function checks
Note
The function checks must be carried out prior to each
surgical application of the unit and after each intraop-
erative battery change!
Note
Only use the device if it is in perfect condition.
►
Make certain that a battery has been installed.
►
Make certain that the attachment is securely
seated.
►
Check that the tool is seated properly.
►
Briefly run the small drill at maximum speed in
clockwise and counterclockwise mode.
►
Make certain that the rotational direction is correct
in each case.
5.3
Safe operation
Coagulation of patient tissue
or risk of burns for patients
and user through hot product!
WARNING
►
►
►
►
►
Risk of infection from aerosol
formation!
Risk of injuries caused by par-
WARNING
ticles coming loose from the
tool!
►
Do not use products for
acetabulum milling.
Cool the tool during opera-
tion.
Keep product/tool out of
the reach of patients.
Let the product/tool cool
down.
Use a cloth to protect
against burns when chang-
ing the tool.
Use suitable protection
(such as waterproof pro-
tective clothing, face
mask, safety gases, suc-
tion).