Message and Code
Code 1033 Repeating measurement to
ensure accuracy.
Code 1037 Not ready.
Solvent Monitor
Installation Guide
Description and Remedial Action
The solvent monitor repeats a measurement when there was an
error during the previous measurement. A measurement will be
repeated up to three times.
If this error persists, refer to the remedies of the preceding error
messages.
Wait for initial measurements to complete.
Check the previous errors in the Chromeleon Audit Trail and
perform the remedial actions listed in the error messages.