Thermo Scientific Vanquish Installation Manual page 40

Solvent monitor
Hide thumbs Also See for Vanquish:
Table of Contents

Advertisement

6
Troubleshooting
Message and Code
Code 1017 System Interlink connection
timeout.
Code 1024 Warning limit reached.
Code 1025 Error limit reached.
Code 1026 System Interlink connection
lost.
Code 1029 Leak detected in the VSM
tubing or connection.
Code 1031 Recoverable channel error.
Please re-enable the channel.
Code 1032 Non-recoverable channel
error.
Page 40
Description and Remedial Action
The connection to a module via the system interlink has timed
out.
Make sure the system interlink cable is attached to the solvent
monitor and the autosampler/detector and that the system
interlink cable is not damaged.
Check whether the modules show the following statuses on their
respective ePanels:
Pump: ready
¨
Solvent monitor: operable
¨
1. Make sure that there is enough solvent in the solvent reservoir/
enough capacity in the waste container.
2. If necessary, check the warning limit parameter and adapt it.
1. Make sure that there is enough solvent in the solvent reservoir/
enough capacity in the waste container.
2. Check the error limit parameter and adjust it, if necessary.
Make sure the system interlink cable is attached to the solvent
monitor and the autosampler/detector and that the system
interlink cable is not damaged.
Check whether the modules show the following statuses on their
respective ePanels:
Pump: ready
¨
¨
Solvent monitor: operable
The pressure dropped during a measurement indicating a leak.
Make sure the solvent monitor line is correctly inserted into the
solvent monitor fitting (see
Lines
(} page 14)).
Push the thrust ring on the solvent monitor port down firmly
and evenly and pull out the solvent monitor line out of the
solvent monitor fitting and cut the first 10 mm with the capillary
cutter.
Check the solvent monitor line and the dipstick for damage. If
necessary, replace the solvent monitor line.
Check whether the solvent monitor line is discolored, kinked or
otherwise damaged. If necessary, replace the solvent monitor
line.
An error has occurred prior to this message so that the channel
cannot deliver results.
Check the previous errors in the Chromeleon Audit Trail and
perform the remedial actions.
Disable and enable the channel to trigger new measurements.
An error has occurred prior to this message so that the channel
cannot deliver results.
Check previous errors in the Chromeleon Audit Trail and perform
the remedial actions listed in the error messages.
If the error persists, disable the channel and use a different one
for your measurement. If all channels are used, replace the
solvent monitor.