Preparations
Follow these steps
Solvent Monitor
Installation Guide
1. Turn off the pump flow and wait until the system pressure is down
to zero.
1. Unscrew the cap of the solvent reservoir/waste container and
remove the solvent monitor line together with the cap from the
solvent reservoir/waste container.
2. Remove the end cap from the dipstick.
3. Pull the solvent monitor line out of the solvent reservoir cap/waste
container cap.
4. Remove the retaining guide.
5. Remove the line label.
6. Pull the line(s) out of the line sleeve.
7. Push the thrust ring on the solvent monitor port down firmly and
evenly and pull out the solvent monitor line. Do not pull the solvent
monitor line without having pushed down the thrust ring.
8. Remove the line label.
9. Long waste monitor lines: Remove the solvent monitor line from the
tubing guide of all modules above the system base in the system
stack. When removing the solvent monitor line from the tubing
guides, be careful not to pull on other lines in the tubing guides.
10. Unpack the replacement solvent monitor line.
11. To connect a new solvent monitor line, follow the steps in
Connecting the Solvent Monitor Lines
12. To remove any air bubbles from the solvent lines, purge the pump.
5
Maintenance and Service
(} page 14).